循环风消毒机与紫外线灯对空气消毒效果的比较  被引量:6

COMPARISON OF AIR DISINFECTION EFFICACY OF CIRCULATING AIR DISINFECTION MACHINE AND ULTRAVIOLET LAMP

在线阅读下载全文

作  者:姚楚水[1] 古希波[1] 丁兰英[1] 杨玉琴[1] 

机构地区:[1]军事医学科学院微生物流行病研究所,北京100071

出  处:《中国消毒学杂志》2001年第2期111-112,共2页Chinese Journal of Disinfection

摘  要:为比较循环风消毒机与紫外线灯对空气的消毒效果,用FA—1型多级撞击式采样器进行了采样检测。结果, ME—1型空气消毒器开机 40 min,可使空气中细菌总数消亡率>94.86%。进入后 60 min内,因继续消毒,空气中细菌总数可维持在 28~ 180 cfu/m3。紫外线灯照射 60 min.可使空气细菌总数减少≥90.54%,但进入后空气细菌总数很快上升。In order to compare the air disinfection efficacy of circulating air disinfection machine and ultraviolet lamp, multigrade impinger--type sampler type FA--1 was used for sampling. The results showed that operation of ME--l--type air disinfector for 40 min could result in total air bacteria decay rate of ≥94.86%. Within 60 min after the room was occupied, continuous disinfection could maintain the total bacterial count of air at 28-180 cfu/ m3. Irradiation of ultraviolet lamp for 60 min could reduce the total bacterial count of air by≥ 90.54%, but the total bacterial count of air increased rapidly after the room was occupied.

关 键 词:循环风空气消毒器 紫外线 空气消毒 

分 类 号:TU831.5[建筑科学—供热、供燃气、通风及空调工程] R187[医药卫生—流行病学]

 