前置自适应统计迭代重建技术对超低剂量CT图像质量影响的模型研究  被引量:16

Effect of pre-setted adaptive statistical iterative reconstruction Veo on image quality of ultra-low dose CT:phantom study

在线阅读下载全文

作  者:何欣 张景东[1] 杨正汉[1] 张婷婷[1] 徐岩[1] HE Xin;ZHANG Jingdong;YANG Zhenghan;ZHANG Tingting;XU Yan(Department of Radiology ,Beijing Friendship Hospital ,Capital Medical University, Beijing 100050,China;Department of Radiology ,Beijing Obstetrics and Gynecology Hospital, Capital Medical University ,Beijing 100010,China)

机构地区:[1]首都医科大学附属北京友谊医院放射科,北京100050 [2]首都医科大学附属北京妇产医院放射科,北京100010

出  处:《实用放射学杂志》2019年第1期110-113,125,共5页Journal of Practical Radiology

基  金:北京市自然科学基金资助项目(面上项目)(7142041).

摘  要:目的 探讨新一代不同权重前置自适应统计迭代重建技术(ASiR-V)对超低辐射剂量胸部CT图像质量及磨玻璃密度结节(GGN)体积测量准确性的影响。 方法 采用256排CT扫描含有6个GGN的胸部模型,扫描方案为:A组,管电压80 kV,噪声指数(NI)=10;B组,管电压80 kV,NI=20;C组,管电压120 kV,NI=10;D组,管电压120 kV,NI=20。每组ASiR-V权重分别设置为0%、30%、50%、70%、100%。比较不同组别图像的平均噪声值(SD)、CNR及GGN体积测量误差之间的差异。 结果 不同权重比例的ASiR-V时4组SD与CNR组内差异均有统计学意义(P<0.05),但A组与C组的SD和CNR组间差异均无统计学意义(P>0.05);0%ASiR-V、30%ASiR-V、50%ASiR-V时B组与D组的SD差异有统计学意义(P<0.05);不同权重比例的ASiR-V时GGN体积测量的绝对百分误差(APE) 4组的组间差异无统计学意义(P>0.05)。 结论 80 kV,NI 20,前置50%ASiR-V扫描时,可有效降低辐射剂量且保证图像诊断的质量。Objective To investigate the effect of the new generation of pre-setted adaptive statistical iterative reconstruction Veo (ASiR-V)technique in different weight on the chest CT linage quality with ultra-low radiation dose and ground glass nodule(GGN) volume measurement accuracy.Methods A 256-detector CT was used to scan a chest phantom including 6 GGN in group A [tube voltage 80 kV,noise index(NI)10],group B (tube voltage 80 kV,NI 9.0),group C (tube voltage 120 kV,NI 10)and group D (tube voltage 120 kV,NI 20).ASiR-V weight levesi were set as 0%,30%,50%,70% and 100%,respectively in all groups.Also,we compared the &fference between mean noise value(SD),CNR and GGN volume measurement error for different groups.Results There were significant differences in SD and CNR of CT image scanned with different weight ASiR-V between intra-group of four groups(P<0.05). However,there was no significant difference in SD and CNR groups between group A and group G with different weight ASiR-V (P>0.05). The SD difference between group 13 and group D was statistically significant with 0%ASiR-V,30%ASiR-V and 50%ASiR-V t P<0.05 ). There was no significant difference in APE of GGN between four groups with different weight ASiR-V (P>0.05).Conclusion The CT scanning protocol with 80 kV,NI 20 and pre-setted 50~ASiR-V can effectively reduce the radiation dose and ensure the quality of image for diagnosis.

关 键 词:计算机体层成像 辐射剂量 图像质量 

分 类 号:R814.42[医药卫生—影像医学与核医学] R814.49[医药卫生—放射医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象