检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]南京大学 [2]江苏省肿瘤防治研究所
出 处:《原子能科学技术》1991年第1期1-6,共6页Atomic Energy Science and Technology
摘 要:NDZ20医用电子直线加速器,属行波反馈型,采用改变工作频率、相位、磁控管输出功率及束流负载的方法调整能量,电子线能量从5—20MeV分七档,X线有10、15MV两档,并设有AFC,ARC,ADC等自动稳定频率、剂量率和照射野均匀度等控制系统,对一些主要部件达到较为先进的指标。该机运行已近两年,停机率86年与87年分别为2.14%和3.53%,剂量稳定性等一系列指标均达到或优于国家GBW4-81所规定指标,两年来治疗病人1366人,疗效显著。NDZ 20 Medical linear accelerator is a kind of travelling wave feedback mo-del. Its energy can be adjusted by changing frequency, phase, output rf-power ofmagnetron and beam loading. It has seven electron energy ranges between 5 to 20MeV and two energy ranges of x-ray (10 MV and 15 MV). Some advanced tech-niques such as AFC, ARC and ADC are used automatically to stabilize controlsystem for frequency, does rate and flatteness of field. Some major componentshave advanced parameter. The machine has run for two years. The ratio of downtime was 2.14% in 1986 and 3.53% in 1987. All parameters have reached or surp-assed the GBW 4-81 stipulated by the National Standards. In the last two years1366 patients have been treated, the clinical cure effect was notable.