衔铁引起MRI伪影的研究  被引量:7

The Study of the MRI Artifacts Caused by Keepers

在线阅读下载全文

作  者:魏斌[1] 张富强[1] 余强[2] 

机构地区:[1]上海第二医科大学附属第九人民医院口腔医学院口腔修复科,上海2000112 [2]上海第二医科大学附属第九人民医院口腔医学院放射科,上海200011

出  处:《上海口腔医学》2002年第2期138-140,共3页Shanghai Journal of Stomatology

摘  要:目的观察各种衔铁引起MRI伪影的程度。方法实验在0.15Tesla永磁型磁共振机上进行,选择5种不同类型、形状的衔铁,分别进行体外和口腔内实验,对MRI伪影的最大径进行测量和比较。结果薄片形衔铁引起伪影的最大径为95.5mm,桩核形为139.4mm。结论薄片形衔铁引起的MRI伪影范围明显小于桩核形衔铁,其最大径之间存在显著性差异。Objectives The aim of this research was to observe the extension of MRI artifacts caused by keepers. Method five samples were imaged in a water phantom and a living human subject by 0.15 Tesla magnetic resonance devices. The maximum diameters of artifacts were measured and compared. Results The maximal diameters of the artifacts caused by disc-shaped and post-shaped keepers were 95.5mm and 139.4mm.Conclusions Compared with the post-type keeper, the artifacts caused by disc-type keepers diminished significantly, and the difference of the maximal diameter reached statistical significance.

关 键 词:衔铁 磁性固位体 伪影 MRI 

分 类 号:R445.2[医药卫生—影像医学与核医学]

 