检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]广东省深圳市龙华人民医院检验科,518109
出 处:《检验医学与临床》2014年第A01期53-55,共3页Laboratory Medicine and Clinic
摘 要:目的对HA-8160全自动糖化血红蛋白仪进行精密度性能评价。方法按照CLSI EP5-A2文件要求,利用高值和正常两水平质控品对HA-8160全自动糖化血红蛋白仪测定糖化血红蛋白(HbAlc)项目的精密度进行评价。结果HA-8160全自动糖化血红蛋白仪测定糖化血红蛋白(HbAlc)项目高值和正常两水平的批内精密度变异系数分别为0.97%和1.45%;批间精密度变异系数分别为0.62%和0.34%;日间精密度变异系数分别为0.63%和0.56%;总精密度变异系数分别为1.31%和1.59%。结论HA-8160全自动糖化血红蛋白仪测定HbAlC的精密度能够满足临床检测要求。Objective To evaluate the precision of the Automatic Glycohemoglobin Analyzer ADAMSTM Alc HA-8160. Methods Precision study was performed according to CLSI EPS-A2 guidelines,two(high and normal) levels of quality control materials were analyzed. Results For high and normal level quality control materials,the withrun CVs were 0.97% and 1.45% ,the between-run CVs were 0.62% and 0.34% ,the between-day CVs were 0.63% and 0.56% ,and the total CVs were 1.31% and 1.59%, respectively. Conclusion The ADAMSTM Alc HA-8180 HbAlc analyzer showed excellent precision,which could meet the requirements of clinical test.