检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]中国船舶重工集团第七一二研究所,武汉430068
出 处:《能源研究与管理》2014年第3期44-45,66,共3页Energy Research and Management
摘 要:PE隔板在高温下完全灼烧后用(1+4)稀硫酸溶液浸取残渣,采用电感耦合等离子体原子发射光谱法测定浸取液中Ni、Co元素含量。选择测定镍、钻的分析谱线分别为231.604、228.615nm。按所选仪器工作条件进行光谱测定,并制作各元素的工作曲线。镍、钻的检出限(3S)分别为0.004、0.002mgm。对2个样品中的2种元素各测定6次,测定值的相对标准偏差在1.21%~5.86%之间。用标准加入法进行回收试验,测得回收率在96.1%~98.1%之间。PE clapboard was completely burned at high temperature, the residue was leached with sulfuric acid(l+4) ICP-AES was applied to the determination of Ni and Co element in the leaching solution. The analytical spectral lines of 231.604, 228.615 rtm were chosen respectively for determination of Ni and Co. Based on the data, working curves were prepared. Detection limits(3 s) of 0.004, 0.002 mg/L were obtained for Ni, Co respectively. Precision was tested by analyzing 2 samples, giving values of RSD's(n =6) in ranges of 1.21%-5.86%. Recovery was tested by standard addition method, giving values of recovery in the range of 96.1%-98.1%.
关 键 词:电感耦合等离子体原子发射光谱法 PE隔板 镍 钴