Sysmex xs-800i对小儿异型淋巴细胞警示的可信性评价  被引量:1

The evaluation of Atypical lympho displayed by Sysmex xs-800i in children

在线阅读下载全文

作  者:张诗海 王洋 孙丽 

机构地区:[1]安徽省儿童医院医学检验中心,安徽230051

出  处:《安徽卫生职业技术学院学报》2014年第4期109-109,98,共2页Journal of Anhui Health Vocational & Technical College

摘  要:目的:探讨Sysmex xs-800i血细胞分析仪在小儿异型淋巴细胞警示与人工镜检的符合性。方法:在生物显微镜油镜下对Sysmex xs-800i血细胞分析仪在小儿异型淋巴细胞报警异常的560例患儿做血细胞形态学检查,然后与仪器分析做统计比较。结果:血细胞Sysmex xs-800i报警在Q-Flag上0-100、100-200、200-300不同分段上都能检出异型淋巴细胞,各个阶段没有明显差异(P〉0.05)。结论:Q-Flag报警在200-300段检出临床有意义的异型淋巴细胞的可能性更大,但手工镜检也是必要的。Objective: To compare manual Atypical lympho cell morphology analysis in 1000 cases with which were shown abnormal in Sysmex xs-800i hematology analyzer. Methods: The oil microscope was used in the observation of cell morphology in 560 cases which were shown abnormal in hematology analyzer, then the two groups were statistically anlyzed. Results: The Atypical lympho cell can be found in the abnormal stages of 0-100,100~200, 200~300 of the Q-Flag in Sysmex xs-800i hematology analyzer(P 〉0.05) with no significant differences. Conclusion: The manual morphological examination can not be replaced analyzer especially some diseases. The possibility of clinical significance of atypical lymphocyte detected in the 200-300 section of the alarmed Q-Flag is much greater.

关 键 词:血细胞分析仪 异型淋巴细胞 小儿 

分 类 号:R446.1[医药卫生—诊断学]

 