应用Detector729的两种IMRT剂量验证方法比较  被引量:2

Comparison of two dosimetric verification methods of IMRT using Detector729

在线阅读下载全文

作  者:翟贺争[1] 刘辉[2] 李卫星[3] 齐伟华[4] 袁继龙[2] 李明生[2] 阮书州 王小春[1] 

机构地区:[1]中国医学科学院放射医学研究所 天津市放射医学与分子核医学重点实验室,天津300192 [2]中国疾病预防控制中心辐射防护与核安全医学所,北京100088 [3]山西省太原市疾病预防控制中心,030012 [4]中国人民解放军军事医学科学院附属医院,北京100071

出  处:《国际放射医学核医学杂志》2014年第5期285-288,共4页International Journal of Radiation Medicine and Nuclear Medicine

基  金:国家自然科举基金(81272511);中国医学科学院放射医学研究所基金(所开1427)

摘  要:目的:比较、分析Detector729用于IMRT质量控制的两种剂量验证方法。方法用Pinnacle 8.0 m计划系统对13例肿瘤患者做IMRT治疗计划,根据Detector729的CT扫描图像,分别做出实际机架角和0°机架角的验证计划,并在Varian 23EX加速器上实施剂量验证;依据剂量偏差(距离偏差)分别为3%(3 mm)、3%(4 mm)和4%(4 mm)的判定标准,得到两种验证方法的γ验证通过率,采用SAS统计学软件对验证结果进行分析。结果 Detector729两次测量的重复性均<0.3%;应用Detector729对IMRT实际机架角验证3%(3 mm)、3%(4 mm)和4%(4 mm)的平均通过率分别为91.2%、92.4%和95.6%,较0°机架角计划验证通过率分别低了2.2%(t=-2.04,P=0.053)、2.2%(t=-2.02,P=0.055)和1.4%(t=-1.97,P=0.060),两种剂量验证方法的通过率差异无统计学意义。结论 Detector729测量结果可信、准确;实际照射剂量验证和0°机架角剂量验证均与验证计划剂量分布基本一致;建议医疗机构采取符合实际治疗情况的剂量验证方法。Objective To compare and analyze the dose distribution of IMRT between fixed-gantry and unfixed-gantry dosimetric verification. Methods Thirteen patients treated by IMRT were cho-sen randomly for this study. For each case, 2 different plans (fixed-gantry plan and unfixed-gantry plan) were worked out with Pinnacle 8.0 m planning system. The dose verification for each plan was carried out with Varian 23EX by using Detector729 with the octagonal phantom OCTAVIUS. The verification results of the dose distribution between the treatment planning system calculation and the measurement were done by the gamma analysis quantization with dose difference and distance to agreement of 3%(3 mm), 3%(4 mm) and 4%(4 mm) respectively, and the data were analyzed by Statistical Analysis System. Results The repeatability of Detector729 was within 0.3%. According to the gamma analysis , the overall average passing rate of unfixed-gantry verification for the dosimetric deviation of 3%(3 mm), 3%(4 mm)and 4%(4 mm) were 91.2%, 92.4%and 95.6%, which were slightly lower compared with the passing rate of fixed-gantry verification of 2.2%(t=-2.04, P=0.053), 2.2%(t=-2.02, P=0.055) and 1.4%(t=-1.97, P=0.060) respectively, and the difference between the two verification methods was not statistically sig-nificant. Conclusions The Detector729 is a reliable and accurate dosimeter. The two dosimetric verifi-cation methods of IMRT by using Detector729 are efficient and valuable. The radiotherapy department should select suitable and available dosimetric verification method for radiotherapy.

关 键 词:质量控制 放射疗法 调强适形 Detector729 剂量验证 

分 类 号:R142[医药卫生—公共卫生与预防医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象