机构地区:[1]南方医科大学,广东广州510515 [2]广东省医学科学院广东省人民医院放射科,广东广州510080 [3]广东省医学科学院 广东省人民医院放射科,广东广州510080
出 处:《中国医学影像技术》2014年第10期1445-1449,共5页Chinese Journal of Medical Imaging Technology
基 金:国家“十二五”科技支撑计划课题项目(2011BAI11B22);广东省科技支撑计划项目基金(2009B030801257);广东省医学科研基金(2013036)
摘 要:目的探讨迭代重建技术(IDose^4)在降低管电压配合不同对比剂碘流率注射方案下行冠状动脉CT血管成像(CCTA)的可行性。方法对10只实验猪分别重复9次CCTA扫描[管电压分别为120 kV(A_1)、100 kV(A_2)、80 kV(A_3),对比剂碘流率分别约为1300 mgI/s(B_1)、1000 mgI/s(B_2)及750 mgI/s(B_3),两两组合成9组方案];对原始图像分别采用滤波反投影法(FBP)及IDose^4重建,将A_1B_1+FBP重建设为对照组。比较各组图像的主、客观指标。结果在相同条件下,IDose^4重建较FBP重建显著降低图像噪声、提高SNR及CNR(P均d0.001)。随管电压及碘流率的降低,图像噪声增大,SNR、CNR降低。采用IDose4重建,除A_3B_1、A_3B_2、A_3B_3组噪声仍明显增加外(P均<0.05),其余各组噪声、SNR、CNR均与对照组无统计学差异(P均>0.05)。采用FBP重建,A_3B_1、A_3B_2、A_3B_3组及A_2B_3组冠状动脉远段可诊断率较对照组明显减低(P均<0.05),而经IDose^4重建后,除A_3B_3组冠状动脉远段可诊断率仍明显低于对照组(P=0.015)外,余各组均明显改善,与对照组无统计学差异(P均>0.05)。结论采用IDose^4重建结合低管电压及低碘流率行CCTA是可行的。Objective To assess the feasibility of coronary computed tomography angiography (CCTA) at a low tube voltage and low iodine flow rate injection protocol by using iterative reconstruction algorithm. Methods Ten healthy swine underwent repeated CCTA with different combination of tube voltage (A1 , 120 kV; A2,100 kV; A3, 80 kV) and iodine flow rate of contrast media (B1, 1300 mgI/s; B2, 1000 mgI/s; B3, 750 mgI/s), resulting in 9 raw data sets. Each set of raw data was reconstructed with filter back projection (FBP) and iterative reconstruction algorithm (IDose4 ), respectively. Group A1 B1 using FBP reconstruction was set as control group. Subjective and objective indicators of each group were compared. Results IDose4 reduced the image noise, enhanced SNR and CNR significantly in the same protocol as compared to FBP (all P〈0. 001). A reduction in the tube voltage and iodine flow rate of contrast medium caused increasing in image noise and decreasing in SNR and CNR. By using IDose4 reconstruction, except the image noise in group A3 B1 , A3 B2 and group A3B3 was still obviously increased (all P〈0.05), the image noise, SNR and CNR in all the other groups were no significantly different from those in the control group (all P〈0.05). Under FBP reconstruction, the diagnosable rate of distal coronary segments in group A3 BI , A3B2, A3 B3 and group A2B3 was obviously reduced (all P〈0.05). After using IDose4 reconstruction, there was no statistical difference from the control group (all P〈0.05) except the diagnosable rate of distal coronary segments in A3 B3 (P= 0.015). Conclusion Compared to the FBP reconstruction, fDose4 reconstruction can significantly reduce image noise, improve SNR, CNR and subjective image scores.
关 键 词:冠状血管 体层摄影术 X线计算机 管电压 对比剂 迭代重建
分 类 号:R814.42[医药卫生—影像医学与核医学] R543.3[医药卫生—放射医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...