作 者:张超[1,2] 穆学涛[1] 王宏[1] 董悦[1] 董玉茹[1] 钟心[1]
机构地区:[1]武警总医院磁共振科,北京100039 [2]徐州医学院研究生学院,江苏221004
出 处:《放射学实践》2014年第10期1151-1154,共4页Radiologic Practice
基 金:武警总医院基金项目(WZ2012061;WZ2010006);发改办高技(2013)2140号重大灾害医学救援卫星综合应用服务示范基金资助
摘 要:目的:比较不同b值、不同数量扩散梯度方向(NDGDs)3.0T磁共振(MRI)扩散张量成像(DTI)对正常人颈髓成像的影响;探讨颈髓最佳DTI参数。方法:随机选取健康志愿者20例,进行颈髓矢状面DTI扫描,选用b值分别为0、200、400、600、800、1000s/mm2,NDGDs选择6、12、20。分别在不同b值、不同NDGDs条件下,分析图像质量以及颈髓各向异性(FA)值、表观扩散系数(ADC)值。结果:当NDGDs不变(20),随b值增加图像信噪比及质量指数递减(F=28.694、30.272,P<0.05),且不同b值条件下FA、ADC值无组间差异性(F=0.495、0.359,P>0.05)。当b值不变(600s/mm2),随NDGDs增加图像质量评分递增(F=11.927,P<0.05),不同NDGDs条件下FA、ADC值无组间差异性(F=0.08、0.583,P>0.05),但NDGDs为20与6、12分别两两比较,FA值之间差异有统计学意义(P<0.05)。结论:使用西门子3.0T磁共振DTI技术,选取b值为600s/mm2,NDGDs为20的条件下,得到最佳的颈髓DTI图像。Objective:To compare the effects of different b values and different numbers of diffusion gradient direc-tions (NDGDs)of MR diffusion tensor imaging (DTI)with 3T MR scanner for normal cervical spinal cord,and to assess the most optimal DTI parameters.Methods:20 healthy volunteers were recruited randomly,cervical spinal cord sagittal DTI scanning were performed with different b values (0,200,400,600,800,1000s/mm2 )and different NDGDs (6,12,20).The image quality,fractional anisotropy (FA)and apparent diffusion coefficient (ADC)were analyzed with different b values and different NDGDs,respectively.Results:With the constant NDGD as 20,the signal-to-noise ratio (SNR)and image qual-ity index decreased with the increasing of b values (F =28.694,30.272,P 〈0.05),there was no significant difference of FA and ADC between different groups of b value (F =0.495,0.359,P 〉0.05).With the constant b value as 600s/mm2 ,the im-age quality score increased with the increasing of NDGD (F =11.927,P 〈0.05),there was no significant difference of FA and ADC between different groups of NDGD (F =0.08,0.583,P 〉0.05),yet there was significant difference between FA values when comparing NDGD as 20 with that as 6 and 12 (P 〈0.05).Conclusion:The optimal DTI image of cervical spinal cord could be obtained with the parameters including the b value as 600s/mm2 and NDGD as 20 using a Siemens 3.0T MR scanner.
分 类 号:R445.2[医药卫生—影像医学与核医学] R322.81[医药卫生—诊断学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.112