PDH技术中F-P腔特性对鉴频曲线的影响(英文)  被引量:1

Influence of Fabry-Perot cavity on frequency discrimination curve in Pound-Drever-Hall method

在线阅读下载全文

作  者:李海成[1,2] 冯丽爽[1,2] 王俊杰[1,2] 

机构地区:[1]北京航空航天大学惯性科学与技术重点实验室,北京100191 [2]北京航空航天大学精密光机电一体化技术教育部重点实验室,北京100191

出  处:《红外与激光工程》2014年第11期3699-3703,共5页Infrared and Laser Engineering

基  金:国家自然科学基金(61171004)

摘  要:详细介绍和分析了基于Pound-Drever-Hall(PDH)方法的激光器稳频技术。仿真分析了法布里-珀罗(F-P)腔镜面反射率及腔内损耗对鉴频曲线的影响,当反射率为0.2、不考虑腔内损耗时,鉴频曲线线性动态范围和灵敏度分别为9.6 MHz和0.2 MHz-1。与此同时,当腔内损耗为1%时,线性动态范围和灵敏度变化为14.5 MHz和0.05 MHz-1。结果证明了镜面反射率及腔内损耗都会对鉴频曲线产生显著影响。A laser frequency stabilization technique based on Pound﹣Drever﹣Hall (PDH) method was introduced and analyzed in detail. The influences of the Fabry﹣Perot(F- P) mirror's reflectivity and loss of F- P cavity were analyzed by numerical simulation . The theoretical linear dynamic range ( LDR ) and sensitivity were 9.6 MHz and 0.2 MHz-1 without loss respectively if the reflectivity was 0.2. Meanwhile, the LDR was less than 14.5 MHz and the sensitivity preceded 0.05 MHz-1 in a situation considering that the cavity loss was about 1%. The results demonstrate that both the two factors cause significant influence on the LDR and sensitivity of the system.

关 键 词:PDH技术 F- P腔 反射率 腔内损耗 

分 类 号:TN24[电子电信—物理电子学]

 