出 处:《Plant Diseases and Pests》2014年第6期25-28,共4页植物病虫害研究(英文版)
基 金:Sponsored by National Scientific and Technological Support Program(2012BAK25B03-12)
摘 要:[ Objective ] To prepare standard work solution of three pesticides used in the intermediate check of gas chromatograph and evaluate the uncertainty of solution. [ Method ] In view of intermediate check needs of gas chromatograph, common pesticides in pesticide residual test were taken as the study objects, i.e., hexachlorabenzene, malathion and Sulfotep, state pesticide reference materials as the materials, volumetry was applied to prepare standard work solution of pesti- cides. Evenness and stability of the work solution was tested by using gas chromatographs installed with electron capture detector (ECD), flame photometric detec- tor (FPD) and flame ionization detector ( FID), respectively, and after the evaluation of concentration, uncertainty of the work solution was assessed from the per- spectives of raw materials, preparation process, evenness and stability. [ Result ] Mass concentrations of the three standard work solution was 0.1 mg/L, 10.0mg/ L, 100.0 mg/L, respectively, and relative expanded uncertainty 2.640%, 1. 980% and 2.400% ( k = 2). [ Conclusion] The study offers technical support for the intermediate check of laboratory apparatuses.[ Objective ] To prepare standard work solution of three pesticides used in the intermediate check of gas chromatograph and evaluate the uncertainty of solution. [ Method ] In view of intermediate check needs of gas chromatograph, common pesticides in pesticide residual test were taken as the study objects, i.e., hexachlorabenzene, malathion and Sulfotep, state pesticide reference materials as the materials, volumetry was applied to prepare standard work solution of pesti- cides. Evenness and stability of the work solution was tested by using gas chromatographs installed with electron capture detector (ECD), flame photometric detec- tor (FPD) and flame ionization detector ( FID), respectively, and after the evaluation of concentration, uncertainty of the work solution was assessed from the per- spectives of raw materials, preparation process, evenness and stability. [ Result ] Mass concentrations of the three standard work solution was 0.1 mg/L, 10.0mg/ L, 100.0 mg/L, respectively, and relative expanded uncertainty 2.640%, 1. 980% and 2.400% ( k = 2). [ Conclusion] The study offers technical support for the intermediate check of laboratory apparatuses.
关 键 词:Gas chromatograph Intermediate check PESTICIDE Uncertainty
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...