经皮黄疸测试仪在满月小儿的应用价值研究  被引量:1

Research of Transcutaneous Jaundice Dector's Application Value in One-month Old Baby

在线阅读下载全文

作  者:曲志萍[1] 沈国红[1] 韩启军[2] 温天莲[1] 

机构地区:[1]山东大学附属济南市中心医院,山东济南250013 [2]济宁医学院附属医院,山东济宁272000

出  处:《医学与哲学(B)》2015年第1期55-56,共2页Medicine & Philosophy(B)

摘  要:通过对57例满月小儿空腹静脉采血,检测血清总胆红素浓度,同时应用经皮黄疸测试仪测定经皮黄疸指数,探讨经皮黄疸测试仪在满月小儿黄疸测定中的应用价值。发现经皮黄疸测试仪测定胆红素与血清胆红素值呈线性正相关,直线回归方程式Y=1.2783X+1.7381,比较有统计学意义(相关系数r=0.953,P<0.001)。以经皮黄疸测试仪测得值3.27mg/dl为病理性黄疸临界值,其敏感度为84.78%,特异度为81.82%。提示经皮黄疸测试仪在满月小儿中有较高的准确性,能较好地应用于满月小儿的黄疸筛查工作。We determined the total serum bilirubin of hollow intravenous blood and the bilirubin level which was determined by transcutaneous jaundice detector of 57 cases one‐month old baby in order to investigate the value of Transcutaneous jaundice detector in jaundice measuring of one‐month old baby .We found that there was a positive linear correlation between serum bilirubin level with bilirubin level which was determined by transcutaneous jaundice detector . Linear regression equation was Y=1 .2783X+1 .7381 with statistical significance (correlation coefficient r=0 .953 , P〈0 .001) . The sensitivity 84 .78% and specificity 81 .82% would be got with pathological jaundice threshold 3 .27mg/dl which was determined by transcutaneous jaundice detector .Finally we got a conclusion that transcutaneous jaundice detector shows high accuracy in one‐month old baby and it can be employed in jaundice screening among one‐month old babies .

关 键 词:经皮黄疸测试仪 血清胆红素 病理性黄疸 

分 类 号:R446.1[医药卫生—诊断学]

 