作 者:杨星[1] 于晓辉[1] 董玲玲[1] 赵富华[1] 万仁玲[1]
出 处:《中国兽药杂志》2015年第3期43-47,共5页Chinese Journal of Veterinary Drug
摘 要:以磺胺间甲氧嘧啶钠注射液为例,阐述永停滴定法含量测定结果的不确定度评定。通过建立数学模型,分析不确定度来源,对各个不确定度分量进行了评估,计算了扩展不确定度,含量测定结果可表述为(95.3±0.8)%(k=2)。结果表明,重复性测量、校准和温度差异引入的不确定度为测量结果不确定度的主要来源。Taking sulfamonomethoxine sodium injection for example, the uncertainty of content determination by dead-stop titration was evaluated. A mathematical mode of uncertainty evaluation was established and the sources of uncertainty were analyzed and evaluated during the whole determination procedure. Finally the expanded uncertainty was calculated and the result of content determination can be expressed as (95.3±0.8)% (k=2). It showed that the main sources of measurement uncertainty were measurement repeatability, calibration and the laboratory temperature variation.
关 键 词:磺胺间甲氧嘧啶钠注射液 含量测定 不确定度 永停滴定法