检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]东北大学中荷生物医学与信息工程学院,辽宁沈阳110819 [2]辽宁省医疗器械检验所,辽宁沈阳110179
出 处:《东北大学学报(自然科学版)》2015年第5期636-640,共5页Journal of Northeastern University(Natural Science)
基 金:国家科技支撑计划项目(2012BAI22B04);国家自然科学基金资助项目(61372014)
摘 要:总结并分析了现有用于评估颌骨密度的四种最常用的方法及不足,提出了用于口腔锥束CT的基于K2HPO4液体体模的定量CT方法,并说明了具体的测量方法.同时对K2HPO4液体体模进行了均匀性与线性度测试、稳定性测试以及与固体体模的对照测试.测试结果表明在口腔锥束CT下基于K2HPO4液体体模的骨密度测量结果与常规CT下基于固体体模的测量结果等效.同时液体体模制作成本低,配置方便灵活,该方法值得临床推广和使用.The current four most commonly used methods to evaluate the jaw bone mineral density was summarized, and the shortage of these methods was analyzed. Quantitative computed tomography (QCT) method was proposed and used for oral cone beam CT (CBCT) based on K2HPO4 liquid phantoms, and specific measurement method was described. Meanwhile the K2HPO4 liquid phantoms uniformity, the linearity, and the stability were tested, and were contrasted with the solid phantoms. The test results showed that the CBCT bone density measurement results based on K2HPO4 liquid phantoms were equivalent with the results of the conventional CT based on solid phantoms. Meanwhile, in view of the low production costs, convenient and flexible configuration, this method was worthy of clinical promotion and application.
关 键 词:口腔锥束CT 定量CT方法 颌骨密度 K2HPO4液体体模 Ca5OH(PO)4固体体模
分 类 号:R814.42[医药卫生—影像医学与核医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.63