北京市环境中累积剂量测量与瞬时剂量率测量比对分析  被引量:4

Comparison and Analysis between Cumulative Dose and Instantaneous Does Rate of Beijing

在线阅读下载全文

作  者:赵广翠[1] 王毅[1] 

机构地区:[1]北京市辐射安全技术中心,北京100089

出  处:《核电子学与探测技术》2015年第4期339-341,共3页Nuclear Electronics & Detection Technology

摘  要:分别对室内及室外环境进行累积剂量测量和瞬时剂量率测量,并对测量结果进行比对分析。比对结果表明,室内累积剂量测量与瞬时剂量率测量E_n≤1,比对结果为满意;2013年北京市环境累积剂量率与瞬时剂量率测量结果相对偏差<30%,比对结果为满意。本次比对在一定程度上验证了累积剂量测量系统和瞬时剂量测量仪器这两类不同原理的剂量监测仪器测量结果的可靠性。Measurements of the cumulative dose and instantaneous dose rate of indoor and outdoor environment respectively are done and the results are compared. The results indicate that the value of En for the cumulative dose and instantaneous dose rate of indoor is less than or equal to 1, so that the result is satisfactory; The relative deviation of cumulative dose rate and the instantaneous dose rate for Beijing environment in 2013 is less than 30%, so the comparison result is satisfactory too. This comparison, to some extent, verify the reliability of the measurement results based on two different principle for radiactive dose monitoring instruments, one for cumulative dose measurement system, the other for the instantaneous dose measurement instrument.

关 键 词:累积剂量 瞬时剂量率 比对 

分 类 号:TL816.7[核科学技术—核技术及应用]

 