作 者:董慧茹[1,2] 何少茹[2] 郑曼利[2] 陈晓博[2] 钟劲[2] 刘玉梅[2] 孙云霞[2]
机构地区:[1]南方医科大学研究生学院,广州510515 [2]广东省人民医院新生儿科(广东省医学科学院),广州510080
出 处:《岭南心血管病杂志》2015年第2期194-198,共5页South China Journal of Cardiovascular Diseases
基 金:国家"十二五"科技支撑计划项目(2011BAI11B22)
摘 要:目的评估超声心排血量(cardiac output,CO)监测仪(ultrasound cardiac output monitor,USCOM)用于机械通气下早产儿测量CO的准确性。方法本研究为前瞻性观察性研究,同时使用USCOM和多普勒超声心动图(echocardiography,ECHO)测量40例机械通气下早产儿生后24 h内的CO,采用Bland-Altman及直线相关分析比较两种方法测量结果的一致性。结果 USCOM和ECHO测得的左CO结果分别为(372±89)m L/min和(382±92)m L/min,两者差值平均值和一致性区间为(-10.4±210.1)m L/min,平均百分误差为52.2%。USCOM与ECHO测得的右CO结果分别为(604±143)m L/min和(423±93)m L/min,两者差值平均值和一致性区间为(481.6±301)m L/min,平均百分误差为113%。结论 USCOM测量需要机械通气的早产儿左、右CO时与ECHO一致性较差,可能与生后早期循环过渡及呼吸机高参数有关。因此,机械通气下早产儿心功能的评估仍需进一步研究。Objectives To evaluate the accuracy of ultrasonic cardiac output(CO) measured by ultrasound cardiac output monitor(USCOM) in preterm infants with mechanical ventilation. Methods A prospective and observational study was conducted in 40 preterm infants with mechanical ventilation within 24 hours after birth. Cardiac output was measured by USCOM and echocardiography(ECHO). Bland-Altman and linear correlation analysis were performed to confirm the consistency. Results Left ventricular outflow(LVO) measured by USCOM and ECHO were(372 ±89)m L / min and(382 ±92) m L / min respectively; while the right ventricular output(RVO) were(604 ±143) m L / min and(423±93) m L / min respectively. Comparison by two techniques for LVO showed the bias and limits of agreement as(-10.4 ±210.1) m L / min, the average percentage error as 52.2%; for RVO, the bias and limits of agreement as(481.6±301) m L / min, the average percentage error as 113%. Conclusions The consistency of USCOM and ECHO in LVO and RVO is poor in preterm infants with mechanical ventilation; which may be associated with early postpartum transition and high ventilation parameters. Cardiac function assessment in critically ill preterm infants with mechanical ventilation still needs further study.
关 键 词:多普勒超声心动图 心排血量 机械通气 早产儿 超声心排血量监测仪
分 类 号:R540.45[医药卫生—心血管疾病]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.22.194.5