检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:祝荣文[1] 杨亦彬[1] 刘永[1] 尚会强[1] 李列宽[1]
机构地区:[1]遵义医学院附属医院血液净化室,遵义563099
出 处:《中国血液净化》2015年第9期569-571,共3页Chinese Journal of Blood Purification
摘 要:目的建立血液透析用反渗水和透析液内毒素检测制度,探讨透析用反渗水和透析液内毒素的合理限值,降低血液透析治疗风险。方法根据《中国药典》2010版第二部收载的细菌内毒素光度检测方法,即ATI动态试管仪定量法测定透析用反渗水和透析液内毒素含量。结果参照ISO标准,共检测15家医院的透析室,其中14家反渗水内毒素﹤0.25EU/ml,合格率为93.3%。8家透析液内毒素﹤0.50EU/ml,合格率为53.3%。整改后反渗水及透析液合格率达100%。结论加强对透析过程各环节持续质量控制,及时采取干预措施,对提高透析质量,降低血液透析治疗风险是非常有帮助的。Objective To establish system of regular water of hemodialysis and detecting toxin in dialy- sate, and to probe into the reasonable limiting range of water of hemodialysis and detecting toxin in dialysate, in purpose of decreasing the risk of hemodialysis treating. Methods According to Bacterial endotoxin pho- tometric detection method which is included in Chinese Pharmacopoeia(Ch.p) 2010 second edition,experi- menting by adopting AT! dynamic test tube apparatus to quantitive determinate reverse osmosis(RO) and vol- ume of toxin in dialysate(reference to ISOStandard). Results after scrutiny for fifteen hospital in total, it is found that toxin in dialysate of above including fourteen hospital is Below 0.25EU/ml,qualified rate reach 93.3 percent,and eight hospital is below 0.50 EU/ml, qualified rate reaeh53.3pereent.after rectification, rate of RO and dialysate reach 100 percent. Conclusions As a result of strengthening persistent quality control in each progress for hemodialysis, taking intervention timely, and improving quality of hemodialysis, to decrease treating risk of hemodialysis is wholly possible and feasible.
分 类 号:R318.16[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.30