检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]解放军451医院医学工程科,西安710054 [2]解放军451医院血液净化中心,西安710054
出 处:《医疗卫生装备》2015年第9期53-54,68,共3页Chinese Medical Equipment Journal
摘 要:目的 :设计一种血压计检测工作台,有效解决血压计出现汞泄漏后到处流淌以及挥发后对检定人员健康造成伤害的问题。方法:根据汞的流动性和挥发性强的特点进行设计,整个装置由台体、排气罩、导流槽、凸台、导流孔、收集仓和换气扇组成;在排风处配有废气过滤网,其中夹有可与空气中挥发的汞蒸气发生化学反应的硫磺粉。结果:该工作台能很好地减少汞污染,减轻汞对人体的损害。结论:该工作台能安全有效地保证血压计维修、检测工作的展开,值得临床推广应用。Objective To design a sphygmomanometer detecting worktable to prevent the leaked mercury from flowing or doing harm to the staff after volatilization. Methods The worktable was composed of a body, an exhaust hood, a flow-guided groove, a base, a flow-guided hole, a collecting bin, a ventilating fan and etc. A filter screen at the ventilating fan was equipped with the sulphur powder to react with the mercury vapor in the air. Results The worktable could reduce mercury pollution and its damage to the staff. Conclusion The worktable facilitates the maintenance and detection of the sphygmomanometer, and thus is worthy popularizing clinically.
分 类 号:R318.6[医药卫生—生物医学工程] TH776.1[医药卫生—基础医学]