作 者:凌六一[1,2] 林攀攀 黄友锐[1] 谢品华[2,3]
机构地区:[1]安徽理工大学电气与信息工程学院,安徽淮南232001 [2]中国科学院安徽光学精密机械研究所,中国科学院环境光学与技术重点实验室,安徽合肥230031 [3]中国科学技术大学环境科学与光电技术学院,安徽合肥230031
出 处:《光电子.激光》2015年第9期1712-1718,共7页Journal of Optoelectronics·Laser
基 金:国家自然科学基金(41305139,41275038);安徽省自然科学基金(1408085QD75);中国科学院战略性先导科技专项(XDB05040200);中国科学院环境光学与技术重点实验室开放基金(2005DP173065-2013-05)资助项目
摘 要:针对LED光谱容易产生漂移从而影响差分光学吸收谱(DOAS)系统的测量性能,研究了温度和驱动电流对蓝光LED(LZ1-00B205)光谱产生漂移的影响,发现LZ1-00B205中心波长随温度和电流的变化分别为+0.025nm/℃和-0.003 2nm/mA。通过对LZ1-00B205进行恒温控制(误差约为±0.1℃)和恒流驱动(电流波动<1%),有效抑制了LED光谱的波长漂移和光强波动。使用改进后的LZ1-00B205作为长光程DOAS系统光源,对大气NO2进行了监测。当光程为760m时,系统的探测限为4×10-9。实验结果表明,光谱稳定后的LED适于做DOAS系统的测量光源。通过稳定LED光谱改善DOAS反演结果,这种方法不需要反复测量参考光谱,简化了DOAS系统的操作过程,同时提高了测量精度。High brightness light-emitting diode (LED), a new type of luminescent device,is a perfect sub- stitute for xenon lamp used as optical source in differential optical absorption spectroscopy (DOAS) sys- tem. Spectral shift often influences DOAS retrievals. The effects of temperature and working current on spectral shift of a blue LED (LZ1-00B205) are discussed,and it is found that the amplitudes of center wavelength shift along with the two parameters are respectively +0. 025 nm/℃ and -0. 0032 nm/mA. Wavelength shift and intensity fluctuation on LZ1-00B205 are effectively restrained by controlling its temperature constant (the error is about 10.1 ℃) and driving it with a constant current (the current fluctuates within 1%). The blue LED, LZ1-00B205,is used as optical source of long-path DOAS for measuring atmospheric NO2. The detection limit of 4× 10^-9 is obtained for the path length of 760 m in the atmosphere. Atmospheric NO2 measurement results are compared with those measured by a com- mercial long path DOAS and the correlation coefficient is 0.98. The experimental results show that the LED could be available to be light source of long-path DOAS if the emission spectrum is steady. Themethod for improving DOAS retrievals by stabilizing LED spectrum does not need to sample the refer- ence spectra, simplifies the procedure of DOAS system and improves measurement accuracy.
关 键 词:长光程差分光学吸收谱(DOAS)技术 LED 波长漂移 大气NO2监测
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.28