手脚污染测量仪手部探测器β表面活度响应的校准  被引量:3

Beta Surface Activity Response Calibration of Hand Detectors in a Hand-Foot Surface Contamination Monitor

在线阅读下载全文

作  者:王勇[1] 刘倍 牛强 李强[1] 冯梅[1] 商洁[1] 

机构地区:[1]中国辐射防护研究院,太原030006 [2]国核示范电站有限责任公司,山东荣成264300 [3]山东核电有限公司,山东海阳265100

出  处:《辐射防护通讯》2015年第3期12-15,共4页Radiation Protection Bulletin

摘  要:手脚表面污染测量仪用于测量辐射工作人员手部和脚部的放射性污染。为方便校准手部探头的表面活度响应,使校准结果更符合实际测量情况,制作了手部模型和90Sr+90Y薄膜放射源,分析研究了人体组织β散射对测量的影响,同时与使用铝衬底源的校准结果进行了比较分析。Hand-foot surface contamination monitors are usually used in nuclear facilities for radiation protec- tion purpose, which measure surface contamination levels on workers' hands and feet. A hand model and two plastic film sources were made to facilitate the calibration of surface activity response of hand detectors in a monitor of this kind, and give rise to calibration results more in line with actual measurements. The effects of beta backscatter from hand tissue on measurement were investigated. The calibration results with aluminum substrate sources used were employed to perform a comparison in the meantime. Their difference was then ana- lyzed.

关 键 词:手脚污染 表面污染 响应 校准 

分 类 号:TL81[核科学技术—核技术及应用]

 