检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
出 处:《分析仪器》2015年第5期73-76,共4页Analytical Instrumentation
摘 要:采用电感耦合等离子体质谱法(ICP-MS)对烟草样品中的镉含量进行检测,对主要影响测量不确定度的样品称量、水分含量测定、待测液体积、消解液移取体积、测量的重复性、标准工作溶液、标准工作曲线等因素进行了分析,按照JJF 1059.1-2012《测量不确定度评定与表示》的规定进行了合成,并得到最终的扩展不确定度ω(Cd)=(0.6963±0.034)mg/kg。Cadmium in tobacco samples was measured by inductively coupled plasma mass spectrometry(ICP-MS).The impacts of major source of measurement uncertainty were discussed.Uncertainty components such as sample weighing,moisture content,the liquid volume,the pipette volume of the digestion solution,measurement repeatability,the standard work-solution and the standard working curve were analyzed and synthesized according to JJF1059.1-2012 Evaluation and Expression of Uncertainty in Measurement,and the final expanded uncertainty wasω(Cd)=(0.6963±0.034)mg/kg.