ICP-AES法测定高含量有机物中镧系元素的消解方法比较  被引量:2

Comparison of the digestion methods for lanthanide in high content of organic matter by ICP-AES

在线阅读下载全文

作  者:马云[1] 马晓鹏[2] 李旭航[2] 王新强[1] 

机构地区:[1]西安石油大学化学工程学院,西安710065 [2]西安石油大学石油工程学院,西安710065

出  处:《分析试验室》2015年第11期1325-1328,共4页Chinese Journal of Analysis Laboratory

摘  要:采用HNO3消解法、HNO3-双氧水消解法、HNO3-HCl O4消解法处理高含量有机物体系中镧系金属样品,用HK-8100型ICP-AES光谱仪测定消解后的镧系金属离子浓度,比较3种酸消解方法的相对误差率,发现HNO3-双氧水消解法中,10 m L样品中HNO3和双氧水用量为1 m L和2 m L时,误差率最小,消解效果最好。Methods of nitric acid digestion,nitric acid- hydrogen peroxide digestion,nitric acid- perchloric acid digestion were used to deal with samples of lanthanide metal in high organic systemsandthe concentration of lanthanide metal after digestion was measured by HK-8100-type spectrometer ICP-AES. After comparingrelative error rate of three species acid digestion method,we found that if the amount ofnitric acid and hydrogen peroxide was 1m L and 2 m L in 10 m L sample,the error rate nitric acid- hydrogen peroxide digestion method was minimal and the result of digestion was the best.

关 键 词:高含有机物 镧系金属 消解方法 ICP-AES光谱仪 

分 类 号:TE352[石油与天然气工程—油气田开发工程]

 