作 者:谭辉[1] 张宏伟[2] 顾宇参[1] 陈曙光[1] 侯晓广 修雁[1] 石洪成[1]
机构地区:[1]复旦大学附属中山医院核医学科,上海200032 [2]复旦大学附属中山医院普外科,上海200032
出 处:《复旦学报(医学版)》2015年第6期716-721,共6页Fudan University Journal of Medical Sciences
基 金:上海卫生系统先进适宜技术推广项目(2013SY008)~~
摘 要:目的通过与SPECT/CT显像进行比较,探讨乳腺专用伽玛显像(breast specific gamma imaging,BSGI)仪诊断乳腺癌的临床应用价值。方法回顾性研究74例因乳腺肿块或乳头溢液接受手术治疗的患者。所有患者均在术前15天内完成BSGI和SPECT/CT检查,其显像资料与术后病理结果进行对照,分别计算这两种检查方法诊断乳腺癌的灵敏度、特异性、准确性、阳性预测值和阴性预测值,诊断效能比较采用配对Fisher确切概率法,P〈0.05为差异有统计学意义。同时计算这两种检查方法对于癌灶最大径为3-15 mm和〉15 mm诊断的灵敏度,以及诊断腋窝淋巴结转移的灵敏度。结果 BSGI和SPECT/CT显像对乳腺癌诊断的灵敏度、特异性、准确性、阳性预测值、阴性预测值分别为91.8%(45/49)、80.0%(20/25)、87.8%(65/74)、90.0%(45/50)、83.3%(20/24)和77.6%(38/49)、80.0%(20/25)、78.4%(58/74)、88.4%(38/43)及64.5%(20/31),且诊断乳腺癌的效能差异有统计学意义(χ^2=5.44,P=0.039 1)。BSGI显像对于癌灶最大径为3-15 mm和〉15mm诊断的灵敏度分别为80.0%(12/15)和97.1%(33/34),相应的SPECT/CT显像灵敏度为60.0%(9/15)和85.3%(29/34);BSGI和SPECT/CT显像诊断腋窝淋巴结转移的灵敏度分别为18.8%(3/16)和37.5%(6/16)。结论相对于SPECT/CT显像,BSGI提高了乳腺癌诊断的灵敏度,但其诊断腋窝淋巴结转移灵敏度较低。Objective To compare the efficiency of breast specific gamma imaging(BSGI)and SPECT/CT for the diagnosis of breast cancer. Methods Seventy-four patients with palpable breast mass or nipple discharge underwent both BSGI and SPECT/CT within15-day interval before surgery,and the result diagnoses were verified with histological examination in retrospective study.The sensitivity,specificity,accuracy,positive predictive value and negative predictive value of BSGI and SPECT/CT for diagnosing breast carcinoma were calculated respectively.Fisher exact probabilities in 2 × 2 table was applied for statistical analysis,P0.05 regarded as statistical difference.The sensitivity of BSGI and SPECT/CT for diagnosing the maximum diameter of breast cancer lesion in 10-15 mm and more than 15 mm,and for diagnosing axillary metastatic lymph nodes was calculated respectively. Results The sensitivity,specificity,accuracy,positive predictive value and negative predictive value of BSGI and SPECT/CT was 91.8%(45/49),80.0%(20/25),87.8%(65/74),90.0%(45/50),83.3%(20/24)and77.6%(38/49),80.0%(20/25),78.4%(58/74),88.4%(38/43),64.5%(20/31),respectively,and there was significant difference between the two methods(χ2=5.44,P=0.039).And the sensitivity of BSGI for diagnosing the maximum diameter of breast cancer lesion in3-15 mm and more than15 mm was 80.0%(12/15)and 97.1%(33/34)respectively.The sensitivity for axillary metastatic lymph nodes was 18.8%(3/16)by BSGI,while SPECT/CT was 60.0%(9/15)and 85.3%(29/34)respectively,and37.5%(6/16)by SPECT/CT. Conclusions Compared with SPECT/CT,BSGI is a useful adjunct imaging modality with higher sensitivity for the diagnosis of breast cancer.However,it has relatively low sensitivity for diagnosing axillary metastatic lymph node.
分 类 号:R817.4[医药卫生—影像医学与核医学] R737.9[医药卫生—放射医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.15