VM-7衰减校准系统的量程拓展方法研究  被引量:1

Research on Extending Measurement Range of VM-7 Attenuator Calibrating System

在线阅读下载全文

作  者:史永彬[1] 于蒙[1] 

机构地区:[1]中国人民解放军92493部队89分队

出  处:《仪器仪表标准化与计量》2016年第1期39-41,共3页Instrument Standardization & Metrology

摘  要:本文针对VM-7衰减校准系统量程小,难以满足大衰减量计量需求的问题,提出了利用射频替代法拓宽VM-7衰减校准系统测量量程的方案,详细介绍了量程拓展实现过程及原理,经实测数据验证该方案可行,能够为大衰减量的计量测试提供有效解决手段。VM-7 is a high precision attenuator measuring instrument with range less than 90 d B. To meet attenuator's wider calibrating range requirement, a new method using RF substitution on VM-7 is brought forward in this paper. Basing on the microwave fundamental and VM-7 construction principle, analysis and research are introduced. System composition principle, implementation process and uncertainty analysis of measurement data is also described here. The actual test proves that it is a feasible solution to fulfi ll the high value attenuator's calibration.

关 键 词:VM-7 衰减 微波 射频替代 