检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]南昌市疾病预防控制中心质量管理科,江西南昌330038
出 处:《职业与健康》2016年第5期607-608,612,共3页Occupation and Health
摘 要:目的评定原子荧光光度计测砷检出限测定结果的不确定度。方法根据JJF 1059.1-2012《测定不确定度评定和表示》的规定,分析各相对标准不确定度分量,计算合成标准不确定度和扩展不确定度。结果检出限测量结果不确定度报告CL=0.057 ng,U=0.025 ng(k=2)。结论为如何评定原子荧光光度计检出限测量结果的不确定度提供了参考方法。[Objective] To evaluate the uncertainty in determination results of arsenic detection limit by atomic fluorescence spectrophotometer.[Methods]According to the Evaluation and Expression of Uncertainty in Measurement (JJF 1059.1-2012), the relative standard components was analyzed, and the synthetic standard uncertainty and expanded uncertainty were calculated. [Results] The report of detection limit uncertainty showed CL=0.057 ng and U=0.025 ng (k=2).[Conclusion] The study provides the reference method for evaluation of uncertainty of atomic fluorescence photometer detection limit measurement.
关 键 词:原子荧光光度计 砷 检出限 不确定度中国图书资料
分 类 号:R115[医药卫生—公共卫生与预防医学]