EASI与标准12导心电图监测QT及QTd的研究  被引量:2

A Comparsion of EASI 12- lead Electrocardiogram and Standard 12-lead Electrocardiogram in Monitoring QT Interval and QT Dispersion

在线阅读下载全文

作  者:邓颖[1] 吴瑛[1] 张亚荣[2] 张澍[3] 

机构地区:[1]首都医科大学护理学院,100069 [2]首都医科大学宣武医院,100053 [3]中国医学科学院阜外心血管病医院,100037

出  处:《临床心电学杂志》2015年第5期353-355,共3页Journal of Clinical Electrocardiology

摘  要:目的以标准12导联心电图为金标准,评价EASI 12-导联心电监测系统对QT间期及QTd监测的有效性。方法同时记录EASI 12-导联心电图和标准12导联心电图,测量并比较两种心电图各导联的QT间期并计算QTd。结果共采集了216例冠状动脉综合征患者的EASI 12-导联心电图和标准12导联心电图,两种心电图各导联的QT间期具有很好的的相关性(r=0.630~0.843,p〈0.05);各导联QT间期配对t检验差异均无统计学意义(p〉0.05)。EASI 12-导联心电图和标准12导联心电图的QTd配对t检验差异均无统计学意义(p〉0.05)。结论 EASI 12-导联与标准12导联心电图在QT间期及QTd测量上具有一致性,为临床QT及QTd的持续监测提供了依据。Objective Using the standard 12-lead ECG as gold standard , we evaluated the effectiveness of EASI 12- lead ECG in monitoring QT interval and QT dispersion. Methods 216 patients were enrolled, EASI 12- lead ECGs and standard 12-lead ECGs were recorded simultaneously. We measured QT interval and QT dispersion about each lead of two kinds electrocardiograms. Results Each lead of two kinds of ECGs had a good agreement (r = 0.630 - 0.843, p 〈 0.05), and the difference of each lead was not statistically significant (paired t- test results p 〉 0.05). The difference of QT dispersion of two kinds of ECGs (paired t- test results p 〉 0.05) was not statistically significant. Conclusions EASI 12- lead ECGs had a good consistency with standard 12-lead ECGs on measuring QT interval and QT dispersion, and it could be used in continuous cardiac monitoring for QT and QT dispersion,

关 键 词:EASI 12-导联心电图 QT间期 QTD 

分 类 号:R540.41[医药卫生—心血管疾病]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象