Bulk Metallic Glasses: MRI Compatibility and Its Correlation with Magnetic Susceptibility  被引量:1

Bulk Metallic Glasses: MRI Compatibility and Its Correlation with Magnetic Susceptibility

在线阅读下载全文

作  者:Da-Bo Zhou Shao-Ping Wang Shao-Gang Wang Hong-Jun Ai Jian Xu 

机构地区:[1]School of Stomatology,China Medical University [2]Shenyang National Laboratory for Materials Science,Institute of Metal Research,Chinese Academy of Sciences

出  处:《Journal of Materials Science & Technology》2016年第6期496-504,共9页材料科学技术(英文版)

基  金:supported by the National Natural Science Foundation of China under Grant No.51571192

摘  要:Several bulk metallic glasses (BMGs) were selected to in vitro assess their magnetic resonance imaging (MRI) compatibility with agarose gel as a phantom, in terms of the extent of susceptibility artifacts in magnetic resonance image. The investigated metals include the Au49Ags.sPd2.3Cu26.9Si16.3, Zr61Ti2Cu2sA112, Cu50.4Nis.0Ti31Zr13 and Ti47Cu38Zr7.5Fe2.5Sn2Si1Ag2, together with pure titanium (CP-Ti) and Co-28Cr-6Mo alloy (ASTM-F799) for comparison. The artifact extent in MR images was quantitatively characterized according to the total volume in reconstructed 3D images with a series of slices under acquisition by fast spin echo (FSE) sequence and gradient echo (GRE) sequence. As indicated, artifact severity of the BMGs is much less than that of the CoCrMo alloy. The AuAgPdCuSi BMG manifested the smallest arti- fact among the four BMGs, while the TiCuZrFeSnSiAg BMG is comparative to the CP-Ti. The MRI compatibility of BMGs is ranked as a sequence of the Au-, Zr-, Cu- and Ti-based alloys. Dependence of material mag- netic susceptibility on artifact extent is also the case of the BMGs, even though it does not follow a simple linear relationship within a range of △χv = 30-180 ppm. These findings are of interest to reveal that the BMGs are potentially applied in the fields associated with an interventional MRI for MRI-guided surgeries.Several bulk metallic glasses (BMGs) were selected to in vitro assess their magnetic resonance imaging (MRI) compatibility with agarose gel as a phantom, in terms of the extent of susceptibility artifacts in magnetic resonance image. The investigated metals include the Au49Ags.sPd2.3Cu26.9Si16.3, Zr61Ti2Cu2sA112, Cu50.4Nis.0Ti31Zr13 and Ti47Cu38Zr7.5Fe2.5Sn2Si1Ag2, together with pure titanium (CP-Ti) and Co-28Cr-6Mo alloy (ASTM-F799) for comparison. The artifact extent in MR images was quantitatively characterized according to the total volume in reconstructed 3D images with a series of slices under acquisition by fast spin echo (FSE) sequence and gradient echo (GRE) sequence. As indicated, artifact severity of the BMGs is much less than that of the CoCrMo alloy. The AuAgPdCuSi BMG manifested the smallest arti- fact among the four BMGs, while the TiCuZrFeSnSiAg BMG is comparative to the CP-Ti. The MRI compatibility of BMGs is ranked as a sequence of the Au-, Zr-, Cu- and Ti-based alloys. Dependence of material mag- netic susceptibility on artifact extent is also the case of the BMGs, even though it does not follow a simple linear relationship within a range of △χv = 30-180 ppm. These findings are of interest to reveal that the BMGs are potentially applied in the fields associated with an interventional MRI for MRI-guided surgeries.

关 键 词:Metallic glass MRI compatibility Magnetic susceptibility Metal-induced artifacts Biomaterials 

分 类 号:TG139.8[一般工业技术—材料科学与工程] O482.531[金属学及工艺—合金]

 