作 者:刘烨[1] 王硕[1] 张向明[1] 崔蓉[2] 孙涵[1] 李书明[1]
机构地区:[1]北京市朝阳区疾病预防控制中心,北京100021 [2]北京大学医学部,北京100191
出 处:《中国卫生检验杂志》2016年第11期1539-1541,共3页Chinese Journal of Health Laboratory Technology
基 金:首都卫生行业发展专项(2011-7023-02)
摘 要:目的建立生活饮用水中7种常见离子(F^-、Cl^-、SO_4^(2-)、NO_3^--N、Br-、NO_2^--N、PO_4^(3-))和5种主要消毒副产物(ClO_2^-、ClO_3^-、BrO_3^-、DCAA和TCAA)的淋洗液在线发生-离子色谱测定方法。方法用Ion Pac AS19为分离柱,以淋洗液自动发生装置在线产生KOH作淋洗液,通过抑制型电导检测器进行检测。测定条件:梯度洗脱,电导池温度为30℃,流速为1.00 ml/min,进样体积为500μl,抑制器电流为112 m A,柱温为25℃。结果 12种阴离子峰面积与浓度具有良好的线性关系,各组分的相关系数(r)为0.999 9~1.000 0。采用低、中、高3种浓度加标回收,回收率为93.2%~101.0%,相对标准偏差(RSD)≤1.06%。结论该方法操作简单,准确,灵敏,适用于地表水、矿泉水和市政末梢水中12种阴离子的检测。Objective To establish a method for the detection of 7 common anions( F^-,Cl^-,SO_4^(2-),NO_3^--N,Br^-,NO_2^--N,PO_4^(3-)) and 5 disinfection by-products( ClO_2^-,ClO_3^-,BrO_3^-,DCAA and TCAA) in drinking water using ion chromatography with on-line eluent generator. Methods Ion Pac AS19 eluent automatic generator and suppressed conductivity detector were used to produce KOH eluent on-line. Detection was performed by suppressed conductivity detector. Determination conditions: gradient elution; conductivity detection cell temperature was set at 30 ℃; flow rate was 1. 00 ml / min; the sample injection was 500 μl; the electric current was 112 m A; the column temperature was at 25 ℃. Results The 12 anions had a good linear relationship between peak area and concentration,and the correlation coefficient( r) was within 0. 999 9-1. 000 0. At low,medium and high spiked level,the recoveries were within 93. 2%-101. 0%,with the relative standard deviation( RSD) ≤1. 06%. Conclusion This method is simple,accurate and sensitive. It can be used for the detection of 12 anions in surface water,mineral water and municipal pipe water.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.145