探索CT球管故障及对其他配件的损害  被引量:4

To Explore the CT Ball Tube Fault and Damage to Other Accessories

在线阅读下载全文

作  者:谭丹华[1] 

机构地区:[1]广东省江门市开平中医院设备科,开平529300

出  处:《中国医疗器械信息》2016年第6期67-69,共3页China Medical Device Information

摘  要:CT机已成为医院不可缺少的常规诊断设备,可提高医疗诊断水平,正因为CT机技术含量高所以造价也昂贵,其维修费用也比较大,其中主要易耗品就是CT球管,因为它是CT机中的重要部件。本文从CT球管原理出发,分析CT球管故障原因及对其他配件的损害,总结维修经验,目的是节约维修成本,提高经济效益。CT has become indispensable to the hospital routine diagnostic equipment, improve the level of medical diagnosis, because CT technology content high so expensive, the maintenance cost is relatively large, maintenance cost as high as one of the main consumables is CT X-ray tube. This is of the most important parts of CT, the ball from the CT tube principle. Analysis CT bulb tube failure causes and of other accessories damage, summarize the maintenance experience, in order to save the maintenance cost, improve economic benefits

关 键 词:球管 维修 故障 配件 损害 

分 类 号:R814.42[医药卫生—影像医学与核医学]

 