XPS分析绝缘样品中磁透镜对中和效果的影响研究  被引量:1

Effect of Magnetic Lens on Neutralization for Insulating Sample in XPS

在线阅读下载全文

作  者:冯婷[1] 冯根生[1] 祁成林[1] 杨彬 

机构地区:[1]北京科技大学冶金与生态工程学院,北京100083 [2]岛津企业管理(中国)有限公司,辽宁沈阳110001

出  处:《分析测试技术与仪器》2016年第3期159-164,共6页Analysis and Testing Technology and Instruments

摘  要:以绝缘样品聚对苯二甲酸乙二醇酯(PET)为X射线光电子能谱(XPS)分析测试对象,探讨了使用不同的X射线源作为激发源时,磁透镜对荷电中和系统中和效果的影响.试验结果表明,当使用单色化Al Kα为激发源时,磁透镜对荷电中和器的中和效果没有显著影响.而使用非单色化双阳极Mg Kα为激发源,磁透镜开启时才能得到良好的荷电中和效果,磁透镜关闭时所测谱峰会因荷电效应而发生峰形畸变.The charge compensation plays an important role in the X-ray photoelectron spectroscopy (XPS) analysis for theinsulating sample. According to the different X-ray sources used in XPS, the magnetic lens had different effect on thecharge compensation for the insulating sample PET. The results showed that, magnetic lens has no obvious effect on thecharge compensation when using a monochromatic Al Kα X-ray Source. When a non-monochromatic dual anode Mg KαX-ray Source is used, good results in the use of magnetic lens are obtained.

关 键 词:X射线光电子能谱 绝缘样品 磁透镜 荷电中和 X射线源 

分 类 号:TH838[机械工程—仪器科学与技术]

 