新型自动比色法测定饮用水中铝含量的不确定度评定  被引量:3

Evaluation of measurement uncertainty of aluminum in drinking water using a new automatic colorimetric method

在线阅读下载全文

作  者:王丽[1] 岳银玲[1] 高圣华[1] 邢方潇[1] 钱乐[1] 阮建明 张岚 

机构地区:[1]中国疾病预防控制中心环境与健康相关产品安全所,北京100050 [2]深圳市清时捷科技有限公司,广东深圳518109

出  处:《中国卫生监督杂志》2016年第5期418-425,共8页Chinese Journal of Health Inspection

基  金:"十二五"国家科技支撑计划重点项目(2006BAJ25B03)

摘  要:目的对饮用水中铝的新型自动比色法进行不确定度评定。方法依据JJF1059.1-2012《测量不确定度评定与表示》,建立数学模型,分析和识别不确定度来源,运用不确定度传播律对新型自动比色法测定饮用水中铝含量的不确定度进行评定。结果用新型自动比色法测定饮用水中铝含量,测量结果为0.111mg/L,合成标准不确定度为0.003mg/L。取包含因子k=2,置信水平约为95%,扩展不确定度为0.006mg/L。结论配制标准系列溶液产生的不确定度分量最大,其次是线性回归,取样体积带来的不确定度分量最小。提示实际工作中,避免使用较小容量器皿配制标准系列溶液可大幅减小测量结果的不确定度。Objective To evaluate measurement uncertainty of aluminum in drinking water determined by a new automatic colorimetric method. Methods According to Evaluation and Expression of Uncertainty in Measurement (JJF1059.1-2012), the mathematical model is established, and uncertainty sources are analyzed and identified. The measurement uncertainty of aluminum in drinking water determined by the new automatic colorimetric method is evaluated using law of propagation of uncertainty. Results The concentration of aluminum in drinking water sample determined by the new automatic colorimetric method was 0.11 lmg/L.The combined standard uncertainty was 0.003 mg/L. The expanded uncertainty is 0.006mg/L, using a coverage factor k=2 which gives a level of confidence of approximately 95%. Conclusion The biggest contribution to uncertainty components is preparation of standard solution series, followed by linear regression, and the contribution of sample volume is smallest. In practical work, the t uncertainty of resuhs can be significantly decreased by avoiding the use of small capacity vessels for preparation of standard solution series

关 键 词:  自动比色 测量不确定度 

分 类 号:R123[医药卫生—环境卫生学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象