摘 要:依据水中锶-90放射化学分析方法(GB 6766-1986),分析了水中^(90)Sr计算公式中的有关参数和分析过程中的影响因素,确定了水中^(90)Sr测量的不确定度分量,并对这些分量进行评定。给出水中^(90)Sr分析不确定度主要是由样品测量计数和仪器探测效率引入的,可以通过适度地增加测量时间来减小样品计数不确定度,但测量时间不宜超过24 h。:According to Radiochemical analysis of Strontium -90 in water ( GB 6766 - 1986), the parameters of 90Sr ealeulation formula and the affecting factor in analysis process are analyzed in this paper; at last, we confirm and evaluate the relevant uncertainty components of Strontium -90 measurement in water. It gives that Detection Uncertainty of 90Sr in water mainly comes from the sample measurement count and the instrument de- tection efficiency. In order to reduce the sample measurement count uncertainty, we ean increase the measuring time appropriately, but it can not be longer than 24 hours.