布鲁菌实时荧光定量聚合酶链式反应试剂盒临床样本检测探索  

Brucella real-time quantitative PCR kit for rapid detection of clinical samples

在线阅读下载全文

作  者:王季秋[1] 赵欣[1] 王照[1] 陈雪娇[1] 王仲秋 张少先[1] 张晓晨[1] 孙玉侠 

机构地区:[1]中国疾病预防控制中心鼠疫布氏菌病预防控制基地布病研究室,白城137000 [2]吉林省通榆县第一人民医院内三科,通榆137200 [3]吉林省白城市高等医学专科学校药学系,白城137000

出  处:《中华地方病学杂志》2016年第11期817-820,共4页Chinese Journal of Endemiology

摘  要:目的采用布鲁菌实时荧光定量聚合酶链式反应检测方法,探讨布鲁菌的早期、快速检测,为临床医生对布鲁菌感染的诊断提供指导。方法根据布鲁菌表面蛋白31基因序列设计引物和探针,建立布鲁菌实时荧光定量聚合酶链式反应检测方法,并验证反应体系的特异性(布鲁菌标准株16M、544A、1330S、菌苗株104M,参考菌株小肠结肠炎耶尔森菌O:9、大肠埃希菌O:157)、重复性(4种不同浓度的标准品),分析反应体系的最低检测限。结果所建立的检测方法仅能对布鲁菌进行特异性扩增:反应体系中不同底物浓度与循环(Ct)值之间呈良好的线性关系(Y=-2.77X+43.19,R2=0.9948),反应体系的最低检测限为1×103 copies/ml;对布鲁菌标准株16M、544A、1330S与104M的检测结果为阳性,0:9型小肠结肠炎耶尔森菌、0:157大肠杆菌的检测结果为阴性;4种不同浓度的标准品和阴性质控品进行4次重复检测。变异系数分别为0.15%、0.16%、0.23%、0.18%。结论利用实时荧光定量聚合酶链式反应检测方法,可快速、准确地检出布鲁菌,对临床布鲁菌病的诊断有指导意义。Objective To establish a real-time fluorescent quantitative polymerase chain reaction assay system for rapid and accurate detection of Brucella and to provide a guidance for clinical doctors in making diagnosis of Brucella infection. Methods A real-time polymerase chain reaction assay was developed for detection of all species and bivors of BruceUa. According to the sequences of Brucella surface.protein gene 31, a probe and two primers were designed. The specificity (Brucella standard strains 16M, 544A, 1330S, vaccine strains 104M, reference strains Yersinia enterocolitica 0 : 9, E. coli O : 157) and repeatability (4 different concentrations of standard) were verified and the lowest limit of detection of the reaction system was analyzed. Results The established method can be used for specific amplification of BruceUa, the different substrate concentrations of the reaction system showed good linearity with the Ct value (Y = - 2.77X + 43.19, R2 = 0.994 8), the lowest limit of detection of reaction system was 1 x 103 copies/ml, the test results of Brucella standard strains 16M, 544A, 1330S and 104M were positive, test results of Yersinia enterocolitica O : 9 and E. coli O : 157 were negative, 4 different concentrations of standard and negative quality control were tested 4 times, and the variation coefficients were 0.15%, 0.16%, 0.23% and 0.18%, respectively. Conclusion A Brucella real-time fluorescent quantitative polymerase chain reaction assay system is established successfully, which could be used to identify Brucella rapidly and accurately.

关 键 词:布鲁菌 感染 荧光 实时定量PCR 

分 类 号:R440[医药卫生—诊断学] R516.7[医药卫生—临床医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象