脉搏血氧波形参数正常人群参考值范围的制定  被引量:6

Reference value range of pulse oximetry plethysmographic waveform parameters in the normal adults

在线阅读下载全文

作  者:李晨[1] 徐军[2] 韩飞 张晓毳 付阳阳[2] 郑亮亮[2] 景道远 曾瑞峰[5] 刘小禾[1] 丁邦晗[5] 田英平[6] 柴艳芬[1] 卢中秋[7] 朱华栋[2] 于学忠[2] 

机构地区:[1]天津医科大学总医院急诊医学科,天津300052 [2]中国医学科学院北京协和医院急诊科,北京100730 [3]迈瑞生物医疗电子股份有限公司,深圳518057 [4]金华市中心医院急诊科,浙江省金华321000 [5]广东省中医院急诊科,广州510120 [6]河北医科大学第二医院急诊科,石家庄050000 [7]温州医科大学附属第一医院急诊科,浙江省温州325000

出  处:《中华急诊医学杂志》2016年第12期1294-1300,共7页Chinese Journal of Emergency Medicine

基  金:国家临床重点专科建设项目(2012-650);国家卫生计生委公益性行业科研专项项目(201502019);睿意急诊医学研究专项基金(R2015014)

摘  要:目的脉搏血氧饱和度监测目前广泛应用于临床。近年来研究显示脉搏血氧波形(pulse oximetry plethysmographic wavefoiTn,POP)能实时反应患者循环状态改变。本研究将POP参数化,通过观察正常人群POP参数的分布,制定其正常参考值范围。方法多中心前瞻性描述性研究。以全国7个城市的健康成人1019人为研究对象,记录性别、年龄、身高、体质量,采集其室温平静状态下脉搏血氧数据,应用基于MATLAB2012a专用软件进行脉搏血氧波形参数及血流灌注指数提取,并制定脉搏血氧波形参数[脉搏血氧波形幅度参数(the amplitude of POP,Amp)、脉搏血氧波形曲线下面积参数(the area under the curve of POP,AUC)]正常参考值范围。结果正常成人脉搏血氧波形参数在男性和女性之间有明显差异。95%置信区间下正常人群Amp正常参考值为(104.8—2298.7)PVA;AUC正常参考值为(3265.8—6028.5)PVPG。男性Amp正常参考值为(129.4—2433.6)PVA;AUC正常参考值为(3319.0—5862.2)PVPG。女性Amp正常参考值为(89.5—2138.2)PVA;AUC正常参考值为(3163.9~5929.9)PVPG。结论脉搏血氧波形参数包括脉搏血氧波形幅度参数和脉搏血氧波形面积参数,在正常成人中有一定的参考范围。Objective To Pulse oximetry saturation has been wildly used clinically. It has been reported that pulse oximetry plethysmographic waveform (POP) reflected the peripheral tissue perfusion. In this study, we parameterized POP, observed the value of POP parameters in normal adults, and established the normal reference value range. Methods A multi-center prospective descriptive study. Total of 1 019 adult volunteers with normovolemia from 7 cities were enrolled in this study. Sex, age, height, weight and pulse oximetry data in awake and spontaneous breathing under in quiet conditions in the room temperature were collected. POP parameters and perfusion index were analyzed using MATLAB 2012a software. The normal reference value ranges of POP parameters, including the amplitude of POP (Amp) and the area under the curve of POP (AUC) , were formulated. Results Statistical differences of POP parameters were detected between men and women in the normal adult. The 95% confidence reference value of POP parameters in normal population was as follows: Amp ( 104. 8-2298. 7) PVA and AUC (3265. 8-6028. 5) PVPG in total, Amp ( 129. 4-2433. 6 ) PVA and AUC ( 3319.0-5862. 2 ) PVPG in male ; Amp ( 89. 5- 2138. 2) PVA and AUC (3163.9-5929. 9) PVPG in female. Conclusions POP, including the amplitude of POP (Amp) and the area under the curve of POP (AUC), had normal reference value ranges in normal adults.

关 键 词:脉搏血氧 波形 幅度 曲线下面积 外周循环 正常人群 参考值 多中心 

分 类 号:R459.7[医药卫生—急诊医学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象