直接测量法测定CZT探测器漏电流与结电容  

Realization of the Direct Method of Measurement of the Leakage Current and the Junction Capacitance for CdZnTe Detector

在线阅读下载全文

作  者:李元东[1] 曾国强[1] 葛良全[1] 谭承君[1] 徐亚东[2] 

机构地区:[1]成都理工大学地学核技术四川省重点实验室,成都610051 [2]西北工业大学航空学院,西安710072

出  处:《核电子学与探测技术》2016年第7期747-749,754,共4页Nuclear Electronics & Detection Technology

基  金:国家自然科学基金(41474159);四川省科技计划项目(2015JQO035)资助

摘  要:为弥补CZT探测器漏电流与结电容间接测量法的不足,采用直接测量法重新设计了测试方案。通过对CZT探测器的测试,给出了更精确的漏电流与结电容的数值,并将之与间接测量法的数据作了相应的对比,为CZT探测器的应用提供了更可靠的依据。In order to make up for the deficiency of the indirect testing methods of the leakage current and the junction capacitance of CdZnTe (CZT) detector, this article adopts direct testing methods of measurement to redesign testing solution. Through the test of CdZnTe detector, we get more accurate values of the leakage current and the junction capacitance and compare them with the values of indirect methods, thus providing more reliable reference for its application in the field of radiation detection.

关 键 词:CZT探测器 直接测量法 漏电流 结电容 

分 类 号:TL814[核科学技术—核技术及应用]

 