基于超声波的相对血容量检测中传播时间影响因素分析  

Influence Factors on Analyzing Transmission Time of Relative Blood Volume Based on Ultrasonic

在线阅读下载全文

作  者:蒋顺忠[1] 秦剑[1] 王彩玲 罗燕斌 易薇[3] 

机构地区:[1]广州大学机械与电气工程学院,广州市510006 [2]广州市信臣科技有限公司,广州市510440 [3]南方医科大学南方医院,广州市510515

出  处:《中国医疗器械杂志》2017年第1期5-8,共4页Chinese Journal of Medical Instrumentation

基  金:广州市属高校科研项目(1201420935)

摘  要:基于超声波的相对血容量检测可用于血液透析患者干体重的评估,相对血容量计算结果来自于超声波在动脉壶中传播时间的测量,而传播时间测量的准确性直接关系到相对血容量结果的可靠性。影响传播时间的因素是多方面的,该文从患者本身、检测装置和外界因素等方面进行了分析论证,并提出了适当的应对措施。Ultrasound-based measurement of relative blood volume can be used to assess patient's dry weight during hemodialysis. The results of relative blood volume were calculated from the ultrasonic transmission time measurement in the arteriol pot, and the accuracy of transmission time measurement is directly related to the reliability of the results of relative blood volume. There are various factors which influence the travel time, this article analyzed patients themselves, measuring device and the external factors, and advised appropriate counter measures.

关 键 词:超声波 相对血容量 传播时间 影响因素 

分 类 号:R318.6[医药卫生—生物医学工程]

 