硅光电管-闪烁体探测器γ谱仪研制  被引量:2

Development of Gamma-Ray Spectrometer based on SiPM-Scintillation Detector

在线阅读下载全文

作  者:刘翎箭 许浒 陈祥磊 郭智荣 屈国普[2] 

机构地区:[1]武汉第二船舶研究设计所,武汉430064 [2]南华大学核科学与技术学院,湖南衡阳421001

出  处:《核电子学与探测技术》2016年第10期1049-1052,共4页Nuclear Electronics & Detection Technology

摘  要:实验研制了硅光电管-闪烁体探测器γ谱仪。该γ谱仪用硅光电倍增管代替普通光电倍增管作为闪烁体探测器的光学读出端,配置闪烁体探测器,构成新型γ谱仪。测试结果表明:新型γ谱仪随温度漂移变化程度大;能量线性较好,线性相关度R为0.9987;配置LaBr_3:10%Ce^(3+)晶体,其能量分辨率为4.3%~4.9%;配置NaI(TI)晶体,其能量分辨率为8.4%。We developed a new gamma- ray spectrorneter based on SiPM -scintillation detector that takes sili- con photomultiplier as light readout instead of photomultiplier. The experimental results show that gamma energy spectrum is affected by temperature. The relative degree of linearity R is 0. 9987 and energy resolution of 662keV γ- ray is 4.3% ~4.9% when scintillation crystal is LaBr3 :10% Ce3+. The energy resolution is 8.4% for γ- ray with the energy of 662 keV when scintillation crystal is NaI (TI). The measured energy resolution is better when emission area of the scintillation crystal is less than the effective photon sensitive area of SiPM.

关 键 词:闪烁体探测器 硅光电倍增管(Si PM) 温度响应 线性 能量分辨率 

分 类 号:TL812.1[核科学技术—核技术及应用]

 