检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]武汉第二船舶研究设计所,武汉430064 [2]南华大学核科学与技术学院,湖南衡阳421001
出 处:《核电子学与探测技术》2016年第10期1049-1052,共4页Nuclear Electronics & Detection Technology
摘 要:实验研制了硅光电管-闪烁体探测器γ谱仪。该γ谱仪用硅光电倍增管代替普通光电倍增管作为闪烁体探测器的光学读出端,配置闪烁体探测器,构成新型γ谱仪。测试结果表明:新型γ谱仪随温度漂移变化程度大;能量线性较好,线性相关度R为0.9987;配置LaBr_3:10%Ce^(3+)晶体,其能量分辨率为4.3%~4.9%;配置NaI(TI)晶体,其能量分辨率为8.4%。We developed a new gamma- ray spectrorneter based on SiPM -scintillation detector that takes sili- con photomultiplier as light readout instead of photomultiplier. The experimental results show that gamma energy spectrum is affected by temperature. The relative degree of linearity R is 0. 9987 and energy resolution of 662keV γ- ray is 4.3% ~4.9% when scintillation crystal is LaBr3 :10% Ce3+. The energy resolution is 8.4% for γ- ray with the energy of 662 keV when scintillation crystal is NaI (TI). The measured energy resolution is better when emission area of the scintillation crystal is less than the effective photon sensitive area of SiPM.
关 键 词:闪烁体探测器 硅光电倍增管(Si PM) 温度响应 线性 能量分辨率
分 类 号:TL812.1[核科学技术—核技术及应用]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.191