机构地区:[1]南阳市中心医院,郑州大学附属南阳医院,河南省南阳市473000 [2]河南省淅川县人民医院,河南省南阳市474450 [3]解放军兰州军区兰州总医院全军骨科中心,甘肃省兰州市730050 [4]兰州大学第二医院骨科,甘肃省兰州市730030
出 处:《中国组织工程研究》2017年第7期1143-1148,共6页Chinese Journal of Tissue Engineering Research
基 金:国家自然科学基金面上项目(81371983)~~
摘 要:背景:全髋关节置换后远期出现的关节无菌性松动直接影响人工关节使用寿命和患者满意度。导致松动的原因在于人工关节磨损产生的颗粒诱导假体周围骨溶解所致,而髋臼内衬磨损产生的聚乙烯颗粒是主要因素。目的:系统评价全髋关节置换后交联聚乙烯和普通聚乙烯髋臼内衬磨损和骨溶解松动情况的差异。方法:按照Cochrane协作网制订的检索策略进行检索,计算机检索截止至2015年12月Medline、EMbase、中国生物医学文献数据库、中国期刊全文数据库、中文科技期刊全文数据库、万方数据库及Cochrane图书馆(2011年第3期)。手工检索相关的中外骨科杂志和学术会议论文汇编。纳入交联聚乙烯和普通聚乙烯内衬全髋关节置换后聚乙烯磨损情况的随机对照试验11篇,对符合纳入标准的研究用RevM an 5.0软件进行Meta分析。结果与结论:(1)纳入随机对照试验11篇,共952例患者;(2)Meta分析结果显示,术后平均5年交联聚乙烯髋臼内衬线性磨损率显著低于传统聚乙烯髋臼[MD=-0.07,CI(-0.09,-0.05),I^2=93%,P<0.000 01],由于异质性很大(I2=93%),行敏感性分析,去掉3个研究后异质性下降[MD=-0.06,95%CI(-0.07,-0.04),I2=39%,P<0.000 01];(3)术后平均5年骨溶解松动Meta分析显示,交联聚乙烯内衬组的骨溶解松动率显著低于普通聚乙烯内衬组[RR=0.39,95%CI(0.27,0.57),I2=0%,P<0.000 01];(4)结果提示,全髋关节置换交联聚乙烯内衬髋臼较普通聚乙烯更耐磨,近期磨损率更低,磨损碎少,引起骨溶解松动的概率越低。但由于搜集的随机对照试验研究平均随访时间5年(1.8-8.0年),长期随访结果不得而知。鉴于纳入研究在方法学方面的局限性,尚需开展多中心、大样本、高质量的随机对照试验进一步论证交联聚乙烯的疗效和安全性。BACKGROUND: Aseptic loosening occurs after long-term total hip replacement, which directly affects the service life and prospective efficacy of artificial joints. The particles produced by artificial joint wear lead to the surrounding bone dissolved, further cause loosening, among which, polyethylene particles because of acetabular liner wear stand out. OBJECTIVE: To systematically assess the acetabular liner wear, loosening and osteolysis caused by cross-linked polyethylene or conventional polyethylene after total hip arthroplasty. METHODS: A computer-based research of Medline, EMbase, CBM, CNKI, Cq Vip, Wan Fang databases before December 2015 and Cochrane(3rd issue, 2011) was performed in accordance with the retrieval strategy made by Cochrane collaboration. A manual retrieval of related bone journals and conference papers was conducted. Eleven randomized controlled trials about the wear caused by cross-linked polyethylene or conventional polyethylene after total hip arthroplasty were enrolled based on inclusion criteria, followed by a Meta-analysis using Rev Man 5.0 software. RESULTS AND CONCLUSION:(1) Eleven randomized controlled trials involving 952 patients were included.(2) Meta-analysis showed that the acetabular liner wear rate of cross-linked polyethylene was significantly lower than that of conventional polyethylene at 5 years postoperatively [MD=-0.07, CI(-0.09,-0.05), I2=93%, P〈 0.000 01]; the large heterogeneity was decreased [MD=-0.06, 95%CI(-0.07,-0.04), I2=39%, P 〈0.000 01] after three research removed through sensitivity analysis.(3) The osteolysis rate in the cross-linked polyethylene group was significantly lower than that in the conventional polyethylene group [RR=0.39, 95%CI(0.27, 0.57), I2=0%, P 〈0.000 01].(4) These results suggest that the cross-linked polyethylene liners exhibit reduced radiological wear and osteolysis, but the mean follow-up of 5 years(1.8 to 8.0) cannot meet the long-term requirements. Therefore, multi-central, large sam
关 键 词:关节成形术 置换 髋 假体植入 Meta分析 组织工程 骨科植入物 人工假体 全髋关节置换 传统聚乙烯 交联聚乙烯 磨损率 系统评价 国家自然科学基金
分 类 号:R318[医药卫生—生物医学工程]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...