作 者:闫书雄[1,2] 朱辉 莫婷 高伟[1,2] 黄正旭[1,2]
机构地区:[1]暨南大学质谱仪器与大气环境研究所,广东广州510632 [2]广东省大气污染在线源解析系统工程技术研究中心,广东广州510632 [3]昆山禾信质谱技术有限公司,江苏昆山215311 [4]广州禾信仪器股份有限公司,广东广州510530
出 处:《质谱学报》2017年第3期294-301,共8页Journal of Chinese Mass Spectrometry Society
基 金:国家重大科学仪器设备开发专项(2011YQ170067);广东省工程中心建设项目(2015B0903029)资助
摘 要:垂直引入式飞行时间质谱仪(O-TOF MS)是以脉冲的工作方式对样品进行检测,对于离子源产生的连续离子流利用率极低。为了使O-TOF MS更好地匹配离子源,在自制的大气压接口垂直引入式飞行时间质谱仪(API-O-TOF MS)的基础上开发了离子阀技术,将原本连续进入推斥区的离子流调制为与推斥脉冲同步的离子团,提高推斥占空比,进而提高离子利用率。通过进一步研究富集原理以及分析富集参数的影响,确定了离子富集的最佳参数。以PPG-1000为实验对象,在同等条件下,离子信号强度普遍提升1个数量级以上。该结果表明,离子阀技术在O-TOF MS中的应用可有效提高仪器的离子利用率。It is known that continuous ion beams generated by normal ion sources cannot be detected completely by an orthogonal-injection time of flight mass spectrometer(O-TOF MS).A duty cycle is defined as by the ratio of the ions accelerated by a TOF pusher to continuous ion beams introduced to the pusher.Typically,the duty cycle ofthe O-TOF MS is between 5% to 20% depending on the instrument geometry and the m/z value of the ion,because most of the ions are lost while they pass through the pusher during the period among the acceleration pulses of the pusher.To solve the low duty cycle problem,a new technique has been put forward to improve the duty cycle of an O-TOF MS to some degree.The core of the technique relies on an ion valve placed on the skimmer2,or the end-cap of the RF quadrupole,that has the function to transmit or stop the ion beam.By using the ion valve technology,the RF quadrupole is turned into an ion trap,which ejects ions axially into the TOF synchronously with the TOF pusher.A delay between the ejection timing of the RF quadrupole and the push timing of the OTOF MS leads to the mass synchronization.The ions can be detected with a duty cycle of near 100%in a certain mass range and a much higher sensitivity can be realized.It is noted that the synchronous mode is only suitable for the circumstances where recording the complete mass spectrum is not necessary but a higher sensitivity is expected for a certain mass range.The limitation of a mass range is caused by the spatial spread which occurs during the flight from skimmer2 to the pusher,because the lower m/zions would move faster than the higher m/zions in the same acceleration DC field if ions obtained the same kinetic energy.Our future effort will be try to improve a duty cycle in a larger mass range.Here,we presented the simulation results for several ion valve configurations and their comparison with experimental measurements.The experiment results showed that the application of an optimized ion valve could improve the peak intensities of some masse
关 键 词:垂直引入式飞行时间质谱仪(O-TOF MS) 离子利用率 离子阀 占空比
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.145