作 者:朱纯生[1] 赵虹[1] 陈庆容[1] 张晶[1] 林晓锐[1]
机构地区:[1]珠海市中西医结合医院南方医科大学附属珠海医院影像科,广东519000
出 处:《放射学实践》2017年第5期526-528,共3页Radiologic Practice
摘 要:目的:探讨不同附加滤过组合对球管负荷和辐射剂量的影响,旨在合理选择附加滤过组合,既能有效减少辐射剂量又不过分损耗X线球管。方法:以设备自带设定膝关节正位、腰椎正位和胸部正位摄影曝光模式,分别进行不同附加滤过组合[无附加滤过(A组)、2mm铝(B组)、0.1mm铜+2mm铝(C组)、0.2mm铜+2mm铝(D组)]对自制模体进行三组曝光,每组模式的曝光因子(除滤过组合外)相同,每个滤过组合进行5次曝光,记录每次曝光的曝光量(mAs)和剂量面积乘积(DAP)值,并进行统计学分析。结果:随着附加滤过厚度的增加,球管负荷随之增大;辐射剂量(DAP)随之减小;两两比较显示,球馆负荷(mAs值)D组>C组>B组>A组(P<0.001),辐射剂量(DAP值)D组<C组<B组<A组(P<0.001)。分析0.1mm铜+1mm铝与2mm铝的滤过效果,显示0.1mm铜比1mm铝的滤过能力更强(P<0.001)。结论:X线摄影时必须选择附加滤过,以0.2mm铜+2mm铝附加滤过效果最好,可以最大限度减少辐射剂量,但球管损伤也大,在实际工作中可灵活选择。Objective:The purpose of this study was to investigate the effect of different additional filtering on X-ray tube load and radiation dose in order to select a reasonable assembly that could effectively reduce the radiation and not increase excessive wear of X-ray tube. Methods:A-P radiography of knee joint,lumbar,and chest of self-made molds was performed based on the built-in settings of the diagnostic X-ray source device by adding different filtering assemblies,0Al, 2mmAl,0. lmmCu+2mmAl,and 0.2mmCu + 2mmAl with other exposure parameters unchanged. A total of 5 exposures were performed under each assembly. The mAs and DPA value of each exposure were documented for statistical analysis. Results:With the increase of additional filtration thickness the tube load (mAs value) increased by 18. 063 %-88. 385 % and the radiation dose (DAP value) reduced by 11. 600% -55. 407 %. The results of single factor analysis of variance and pairwise comparison showed that the tube load (mAs value) was 0.2mmCu+ lmmAl 〉 0. lmmCu+ 1mmAl 〉 2mmAl〉0AI (P〈0. 001) ;the amount of radiation (DAP value) was 0.2mmCu+ lmmAl 〈 0. 1mmCu+ lmmAl 〈 2mmAl (P 〈0. 001). The result of contrast analysis of 0. 1mmCu +1mmAl and 2mmAl indicated that the 0. lmmCu was better than lmmAl in filter capacity (P 〈 0. 001). Conclusion:Additional filtration must be selected by users in radiography. The assembly of 0.2mmCu + 2mmAl is of the best filtration effect. It can minimize the radiation dose but the wear is also noticeable, which provides as an alternative in practical work.
分 类 号:R814.2[医药卫生—影像医学与核医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.30