Celect滤器回收困难病例的取出技巧  被引量:4

Technique for retrieval of celect filters in difficult cases

在线阅读下载全文

作  者:李兆南 徐燕能 张向琼 郑波 胡伟 马小燕 斯光晏 

机构地区:[1]西南医科大学附属中医医院介入诊疗科,四川泸州646000

出  处:《中国介入影像与治疗学》2017年第6期347-350,共4页Chinese Journal of Interventional Imaging and Therapy

摘  要:目的探讨复杂性回收钩贴壁下腔静脉滤器的取出技巧及应用策略。方法收集因滤器回收钩贴壁、采用Günther Celect配套回收装置无法成功取出Celect下腔静脉滤器9例,采用交换导丝成襻法行下腔静脉滤器取出术。结果 9例患者,最长植入时间为142天,最短37天,平均(88.67±33.85)天。使用交换导丝成襻法8枚滤器成功取出,1例患者因下腔静脉严重弯曲致回收失败,滤器回收成功率为88.89%(8/9)。滤器回收耗时57~162min,平均(69.89±12.12)min。所有滤器取出后形态完整,术后下腔静脉造影复查未见管壁穿孔,对比剂外渗。结论采用交换导丝成襻法可有效提高回收钩贴壁下腔静脉滤器成功回收率,具有一定的临床应用价值。Objective To explore the techniques and strategies for the retrieval of the retractable inferior vena cava filter (IVCF). Methods Celect IVCF retrieval was not removed successfully with Günther Celect recovery device in 9 cases and exchange-wire-loop removal of inferior vena cava filter method were applicated. Results The longest implanted time of 9 patients was 142 days, the shortest implanted time was 37 days, and the average time was (88.67±33.85)days. Eight filters were successfully removed and one failed due to severe bending of inferior vena cava. Filter retrieval rate was 88.89% (8/9). The average retrieval time was (69.89±12.12)min (57-162 min). No perforation of the wall and contrast agent were found in all patients after the inferior vena cava angiography. Conclusion For the retrieval of the hook heavily attached to the IVCF, the retrieval technique of using the exchange wire into the loop method can effectively improve the retrieval rate and has a certain clinical value.

关 键 词:肺栓塞 可回收滤器 导丝 静脉血栓形成 

分 类 号:R654.4[医药卫生—外科学]

 

参考文献:

正在载入数据...

 

二级参考文献:

正在载入数据...

 

耦合文献:

正在载入数据...

 

引证文献:

正在载入数据...

 

二级引证文献:

正在载入数据...

 

同被引文献:

正在载入数据...

 

相关期刊文献:

正在载入数据...

相关的主题
相关的作者对象
相关的机构对象