检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:贾延兵 刘浩 熊愿 王楚珊 洪文侠 钟林森 黄战武 张世杰 JIA Yan-bing LIU Hao XIONG Yuan WANG Chu-shan HONG Wen-xia ZHONG Lin-sen HUANG Zhan-wu ZHANG Shi-jie(a. Department of Physical Therapy b. Guangdong-Hongkong Joint Sports Injury Rehabilitation and Research Center, Guangdong Provincial Work Injury Rehabilitation Hospital, Guangzhou, Guangdong 510440, Chin 2. Rehabili- tation Department, JORU Rehabilitation Hospital, Yixing, Jiangsu 214200, Chin 3. Department of Physical Therapy, Dongguan Rehabilitation Hospital, Dongguan, Guangdong 523119, Chin)
机构地区:[1]广东省工伤康复医院物理治疗科,广东广州市510440 [2]宜兴九如城康复医院康复治疗部,江苏宜兴市214200 [3]广东省工伤康复医院粤港运动损伤康复与研究中心,广东广州市510440 [4]东莞市康复医院物理治疗科,广东东莞市523119
出 处:《中国康复理论与实践》2017年第5期572-574,共3页Chinese Journal of Rehabilitation Theory and Practice
摘 要:目的测试手持型肌力测定仪OE-210测定下肢肌力的信度与效度。方法 2016年3月1日至4月30日,正常青年人38名,由2位治疗师分别使用OE-210进行股四头肌和腘绳肌肌力测定,其中一位治疗师在3 d后重复测试1次。受试者1 d后接受下肢等速肌力测试。计算组内相关系数(ICC),OE-210测试结果与等速测试结果进行Pearson相关分析。结果同一测试者重复测量ICC=0.718~0.924;不同测试者间ICC=0.784~0.870。OE-210测试结果与等速测试峰力矩正显著相关性(P<0.001),其中股四头肌轻中度相关(r=0.270~0.413),腘绳肌高度相关(r=0.582~0.668)。结论 OE-210肌力测定仪测试下肢股四头肌、腘绳肌的信度良好,而确保其使用效度的条件尚需进一步研究。Objective To investigate the reliability and validity of a novel hand-held dynamometer, OE-210, for muscles strength of lower extremities. Methods From March 1st to August 30th, 2016, 38 young adults were tested the muscle strength of quadriceps and hamstrings with OE-210 dynamometer by 2 rators, and were retested by one of the raters three days later. The isokinetic test was also conducted on all the subjects one day afterwards. The intraclass correlation coefficient (ICC) of OE-210 test results and the Pearson's correlation coefficient between results of OE-210 and isokinetic test were calculated. Results The ICC of test-retest were 0.718 to 0.924, and the ICC of inter-rater were 0.784 to 0.870. The correlation between muscle performance measured with 2 tools were significant (P〈0.001), that was light to medium on quadriceps (r=0.270-0.413), and strong on hamstrings (r=0.582-0.668). Conclusion OE-210 dynamometer was reliable for muscle strength measurement on quadriceps and hamstrings, and the conditions for valid application need further research.
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:3.141.42.23