红外热像仪与表面热电偶测量发射率的匹配法  被引量:9

Match method of emissivity measurement based on infrared thermal imager and surface thermometer

在线阅读下载全文

作  者:李文军[1] 徐永达 郑永军[1] 

机构地区:[1]中国计量大学计量测试工程学院,浙江杭州310018

出  处:《中国测试》2017年第6期12-15,共4页China Measurement & Test

基  金:国家质量监督检验检疫总局质检公益性行业科研专项(201410133)

摘  要:使用红外热像仪测量物体温度时,需要知道被测物体的发射率。为测量物体的发射率,提出红外热像仪和表面热电偶组合的匹配法。建立匹配法测量发射率的实验装置,使用该实验装置,利用恒温源对测试样品进行加热,在近环境温度范围内对两种测试样品3M Scotch Super 88型绝缘胶带和3M 1721型绝缘胶带的发射率进行测量。实验结果表明,测量值与文献值一致。该文给出两种样品在不同温度区间的具体发射率值,该发射率值可用做红外热像仪工程测温中的参考值。The temperature measurement with an infrared thermal imager requires the emissivity values of the object. A match method of infrared thermal imager and surface thermometer is given for the measurement of emissivity. The experiment device based on match method is established. With the device, two kinds of insulated rubber tape 3M Scotch Super 88 type and 3M 1721 are heated by a constant temperature source close to the environmental temperature and their emissivity is measured. The experiment results show that measured values are consistent with the literature value. Specific emissivity values of samples are given in different temperature range. The emissivity values can be used as reference data in engineering measurement of infrared thermal imager.

关 键 词:红外热像仪 表面热电偶 发射率 匹配法 

分 类 号:O6-3[理学—化学]

 