脉冲X光机的光源特性参数研究  被引量:1

Study on Characteristic Parameter of Pulsed X-ray Generator

在线阅读下载全文

作  者:司粉妮[1] 李名加[1] 王文川[1] 朱学彬[1] 王栋[1] 徐荣昆[1] 张传飞[1] 

机构地区:[1]中国工程物理研究院核物理与化学研究所,四川绵阳621900

出  处:《原子能科学技术》2017年第7期1311-1315,共5页Atomic Energy Science and Technology

基  金:中国工程物理研究院科学技术发展基金资助项目(2014B0103004;2013A0103003)

摘  要:采用模拟计算和实验测量相结合的方法对瑞典Scandiflash AB公司的脉冲X光机的光源特性参数进行详细研究。基于MCNP5编写蒙特卡罗程序对脉冲X光机的二极管进行建模,并模拟计算X射线的能谱分布;采用热释光剂量计对X光源的剂量进行监测,获得了单个阳极丝寿命期内的剂量数据;用闪烁探测器对脉冲X光机产生的X射线进行测量,获得了X光源的强度分布、脉冲半宽度、抖动规律等。The characteristic parameters of pulsed X-ray generator were studied in this paper.The geometric structure of diode of the X-ray generator and the process of X-ray generation were modeled by MCNP5,and the energy spectrum of X-ray was calculated.Doses of X-ray were measured by luminescence dosimeters in the lifetime of one anode(20shots).Intensity distribution and pulse width(FWHM)of X-ray were measured by the scintillation detector.Jitters of both the X-ray generator and high voltage pulse generator were also measured.

关 键 词:脉冲X光机 MCNP5 闪烁探测器 

分 类 号:TL929[核科学技术—核燃料循环与材料] TL816

 