机构地区:[1]合肥市第二人民医院呼吸内科,安徽合肥233000
出 处:《临床肺科杂志》2017年第10期1788-1791,共4页Journal of Clinical Pulmonary Medicine
摘 要:目的探讨Ⅳ型便携式睡眠监测仪(PDMs)对阻塞性睡眠呼吸暂停低通气综合征(OSAHS)的诊断价值。方法通过门诊筛查打鼾患者102例,使用Ⅳ型PDMs和多导睡眠监测(PSG)同时对每位患者行睡眠监测,记录最低血氧饱和度(LSa O2)、睡眠呼吸紊乱指数(AHI),运用Clopper-pearson法计算诊断符合率,Bland-Altman图行一致性分析,评价其诊断价值。结果两者诊断一致率为97.05%,灵敏度为100.00%,特异度为76.92%,Kappa=0.853(P<0.001);两者对OSAHS分度一致率为82.35%,Kappa=0.742(P<0.001)。作Bland-Altman图分析,结果一致性非常高。结论Ⅳ型PDMs对OSAHS患者有较高诊断价值。Objective To investigate the diagnostic value of type IV portable sleep monitor (PDMs) in pa-tients with obstructive sleep apnea hypopnea syndrome (OSAHS). Methods 102 patients with snoring were given type IV PDMs and polysomnography (PSG) for sleep monitoring. The lowest oxygen saturation ( LSa02) and sleep apnea hypopnea index (AHI) were recorded, and the diagnostic rate was calculated by Clopper-pearson method, and consistency by Bland-Altman chart analysis. Results The diagnostic concordance rate of the two groups was 97. 05% , the sensitivity 100% , the specificity 76. 92% and Kappa =0. 853 ( P 〈0. 001 ). The concordance rate of OSAHS between the two groups was 82. 35% , and Kappa = 0. 742 ( P 〈 0. 001 ). Bland-Altman diagram analysis showed high consistency. Conclusion PDMs has high diagnostic value in patients with OSAHS.