作 者:王诗鸣[1] 王诗啸 杨毅 金瑞[1] 陆云峰[1] WANG Shiming WANG Shixiao YANG Yi JIN Rui LU Yunfeng(Department of Radiology, the First Affiliated Hospital of Chongqing Medical University, Chongqing 400016, China Power Engineering Department College of Energy Power and Mechanical Engineering, North China Electric Power University, Baoding 071003, China Department of Medical hnaging, San Tai County People's Hospital, Santai 621100, China)
机构地区:[1]重庆医科大学附属第一医院放射科,重庆400016 [2]华北电力大学能源动力与机械工程学院动力工程系,河北保定071003 [3]三台县人民医院医学影像科,四川三台621100
出 处:《中国医学物理学杂志》2017年第10期1004-1007,共4页Chinese Journal of Medical Physics
基 金:国家临床重点专科建设项目[国卫办医函(2013)544号];重庆医科大学附属第一医院培育基金(PYJJ2017-29)
摘 要:目的:对比分析数字断层融合(DTS)和数字X线摄影(DR)技术在寰枢椎关节检查中的应用价值。方法:从既往的寰枢椎检查病例中,分别选取50例已行寰枢椎DTS检查和50例行DR寰枢椎张口位检查的患者,调取其影像信息,由两名高年资医师对其进行诊断,并分别对其影像质量进行评分,诊断结果与已明确的临床结果进行对比。结果:DTS组的诊断准确率(96%)明显高于DR组(78%),差异有统计学意义(P<0.05);DTS组的影像质量评分(4.140±1.080)也明显高于DR组(2.800±1.294)。结论:相较于DR技术,DTS技术在寰枢椎的鉴别诊断中具有获取的影像质量高、诊断准确率高、检查过程中对患者的配合度要求低等优势,加之辐射剂量不高、检查费用也未明显增多,因此DTS技术可作为寰枢椎检查的常规技术手段。Objective To assess the value of digital tomosynthesis(DTS) and digital radiology(DR) in atlantoaxial joint examination. Methods Two experienced doctors diagnosed the atlantoaxial joint of 50 patients examined with DTS and 50 patients with DR selected from the previous atlantoaxial joint patients through their related images, and then scored the image quality.The diagnoses were compared with those confirmed clinical results. Results The diagnostic accuracy in DTS group was higher than that in DR group, with statistical significance(96% vs 78%; P〈0.05), and the image quality score in DTS group was also significantly higher than that in DR group [(4.140±1.080) vs(2.800±1.294)]. Conclusion DTS can be used as a conventional technique in the atlantoaxial joint examination. Compared to DR, DTS in the diagnosis of atlantoaxial joint can achieve a higher quality image, a higher diagnostic accuracy and has lower demands for patients, with a low total radiation dose and without increasing the examination charge.
分 类 号:R814[医药卫生—影像医学与核医学]
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在载入数据...
正在链接到云南高校图书馆文献保障联盟下载...
云南高校图书馆联盟文献共享服务平台 版权所有©
您的IP:216.73.216.229