检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
作 者:陈剑[1,2] 吴昊 刘安雯[1] 胡水明[1] 张劲松[1,2]
机构地区:[1]中国科学技术大学,合肥微尺度国家实验室(筹),合肥230026 [2]加州大学河滨分校化学系及空气污染研究中心,加州92521
出 处:《Chinese Journal of Chemical Physics》2017年第5期493-498,I0001,共7页化学物理学报(英文)
摘 要:A two-channel thermal dissociation cavity ring down spectroscopy (CRDS) instrument has been built for in situ, real-time measurement of NO2 and total RNO2 (peroxy nitrates and alkyl nitrates) in ambient air, with a NO2 detection limit of 0.10 ppbv at 1 s. A 6-day long measurement was conducted at urban site of Hefei by using the CRDS instrument with a time resolution of 3 s. A commercial molybdenum converted chemiluminescence (Mo-CL) instrument was also used for comparison. The average RNO2 concentration in the 6 days was measured to be 1.94 ppbv. The Mo-CL instrument overestimated the NO2 concentration by a bias of +1.69 ppbv in average, for the reason that it cannot distinguish RNO2 from NO2. The relative bias could be over 100% during the afternoon hours when NO2 was low but RNO2 was high.
关 键 词:Cavity ring-down spectroscopy Molybdenum converted chemiluminescence Thermal dissociation Organic nitrates