检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]西安科技大学地质与环境学院,陕西西安710054 [2]常州清流环保科技有限公司,江苏常州213144
出 处:《电镀与涂饰》2018年第1期32-35,共4页Electroplating & Finishing
基 金:国家自然科学基金(51278418)
摘 要:考察了增敏剂、还原剂和载流液对氢化物发生-原子荧光光谱法测定电镀废水中铅的影响,确定了以0.2%(体积分数)的100 g/L铁氰化钾溶液作为增敏剂,还原剂为含有0.5%氢氧化钾和1.0%硼氢化钾的溶液,载流液为1.5%(体积分数)的盐酸溶液。该方法的铅检出限为0.045μg/L,加标回收率为98.5%~100.3%。The effects of sensitizer, reducing agent and carrier solution on determination of Pb in electroplating wastewater by hydride generation atomic fluorescence spectrometry were studied. The sensitizer was determined to be a 100 g/L potassium ferricyanide solution with a dosage of 0.2vol.%. The reducing agent is 0.5% KOH + 1.0% KBH4 solution. The carrier solution is composed of 1.5vol.% hydrochloric acid. The method has a detection limit of lead as low as 0.045 μg/L and a recovery ranging from 98.5% to 100.3%.
关 键 词:电镀废水 铅 氢化物发生法 原子荧光光谱法 测定
分 类 号:TQ153[化学工程—电化学工业] O657.39[理学—分析化学]