检索规则说明:AND代表“并且”;OR代表“或者”;NOT代表“不包含”;(注意必须大写,运算符两边需空一格)
检 索 范 例 :范例一: (K=图书馆学 OR K=情报学) AND A=范并思 范例二:J=计算机应用与软件 AND (U=C++ OR U=Basic) NOT M=Visual
机构地区:[1]上海理工大学环境与建筑学院,上海200093
出 处:《广州化工》2018年第3期109-111,148,共4页GuangZhou Chemical Industry
摘 要:研究了将重铬酸盐法步骤与光度滴定法原理相结合滴定水中的COD,从而验证了全自动消解-光度滴定COD仪器研发的可行性。实验结果分析:实验测定滴定终点红褐色溶液的最佳吸光度波长在510 nm处;在510 nm处不同浓度的待测液在到达滴定终点时均有明显的吸光度突变;通过误差分析,验证了光度滴定判断COD终点的方法可行;结合消解器与光度滴定,全自动消解与滴定操作验证了自动消解-光度滴定COD仪器的研发是可行的。The concentration of COD in the water combined the dichromate method with photometric titration method was determined,and the feasibility of research and development of automatic digestion-photometric titration COD measuring instrument was verified. According to the results of research,the conclusions can be made as follows: when the solution got brown which meant that it arrived end-point during titration,the wave length of optimum absorbance of the brown solution was 510 nm. There were obvious increases during titration with different concentration of the samples. It was feasible to use photometric titration method to measure COD in the water after error analysis. This examination demonstrated that it was feasible to combine automatic digestion instrument with photometric titration instrument to be an integrated instrument-automatic digestion-photometric titration COD measuring instrument.
关 键 词:COD 重铬酸盐法 光度滴定法 全自动消解-光度滴定COD仪器
分 类 号:X824[环境科学与工程—环境工程]